10. Error solutions

Error

possible cause

solution

display:"no wire!"

- wire outside measuring range
- incidence of extraneous light
- wire too thin
- wire slag too big

check and correct wire position
use black light protector cap
use wire with 0.2 – 0.3 mm Ø
adjust wire tension

 blinking

accu nearly empty

charge accu

no radio connection to

the ACS-1000 RC

ID is not matching with the ACS-1000

check and change ID

no display of length

position

- out of range
- reflection area too dark
- too much foreign light on the reflection area
- sight glas dirty

divide measurements in sections
use brighter reflector (white)
install light protectors (i.e. window blinds)
clean sight glass

measured values not

plausible

- incidence of extraneous light
- system not compensated
- too much dirt on sensor head
- system focus out of alignment

- accu is loading

use provided black light protector cap
run compensation
send unit for service
check  position  of  machine  slide

(see chapter 8.5)
during  the  loading  process  there  are

larger  voltage  fluctuations,  which  can

distort the measurements

display remains dark

- accu empty or defect
- fuse defect
- unit defect

charge accu resp. have accu changed
change fuse (chapter 12.1)
send unit for service

display non-readable

altered contrast

Readjust contrast

11. Activity to avoid interruptions

1.  If the  unit is  not in  use  cover  the  sensor  head

with the red protector cap to minimize pollution

2.  During  operation  cover  the  sensor  head  with

the  black  protector  cap  to  avoid  strong

exposure to light and pollution

3.  During  measurements  protect  the  ACS-1000

from exposure to direct solar radiation and high

temperature fluctuations

4.  For  safekeeping  store  the  ACS-1000  in  the

provided transport case when not used

11.1 Service address

Hubitron Steuerungstechnik
Hauptstr. 4/1
89611 Obermarchtal / Germany
Tel: 07375/92 20 66
Fax: 07375/92 20 77
Email: [email protected]
Internet: www.Hubitron.de

12. Maintenance and care

For  a  continuous  and  safe  operation  of  the  ACS-

1000 comply with the following steps:

  After use clean the unit  with  a  clean  and  damp

cloth. 

Do not use compressed air!

  Before  each  new  measuring  operation  clean

sight glas with a clean and damp cloth.

  Charge the accu every 6 months, even if not in

use at the time.

  Protect  sensor  head  with  the  provided  red

protector cap from pollution when not in use.

  Store  the  ACS-1000  always  in  the  provided

transport case.

12.1 Change fuse

Should  the  ACS-1000  not  function  anymore,  the

fuse might  be  defect.  Installed  is  a  microfuse  5x20

with 2AT.

  shut off unit

  take off the aluminum blinds of the housing

  loosen the 4 outer screws

  flip cover to the side

  check  and  possibly  change  the  fuse  on  the

board

Summary of Contents for ACS-1000

Page 1: ...ation and arrangement 4 8 1 1 Installation 4 8 1 2 Arrangement 5 8 2 Wire spanning and arranging 5 8 3 Height compensation 6 8 3 1 Adjustment of incline 6 8 4 Length compensation 7 8 4 1 Delete compen...

Page 2: ...t case ACS 1000 1 Protector cap for sensor red 1 Light protector cap black 1 Inbus key SW4 4 Optional equipment 4 1 Remote control ACS 1000 RC With the remote control ACS 1000 RC all relevant function...

Page 3: ...play This display appears for appr 2 seconds when the unit is switched on The SW version and the ID for the remote control is displayed 7 2 2 Main menu On the bottom the assignment of the function key...

Page 4: ...rmation the next instruction will follow back to the compensation menu 7 2 5 Starting menu Graphic display set display data to 0 set display data to absolut switch to text display back to the compensa...

Page 5: ...heet of paper DIN A4 position machine at the right outer stop arrange ACS 1000 that the laser beam remains within the reflection area A sideways deviation can be corrected by turning the ACS 1000 on t...

Page 6: ...ION Now do not move the Z axis anymore move machine slide up until H 4 5 mm and is displayed confirm with NOTE Is the deviation larger 0 1 mm the incline of the sensor head must be corrected with the...

Page 7: ...Is the length compensation not needed the last compensation data must be deleted start length compensation main menu delete data with 8 5 Start measurements Is the ACS 1000 arranged and compensated th...

Page 8: ...esp have accu changed change fuse chapter 12 1 send unit for service display non readable altered contrast Readjust contrast 11 Activity to avoid interruptions 1 If the unit is not in use cover the se...

Page 9: ...Charge goes off the charging process is finished 12 3 Change accu Is the accu running time considerably declining the accu might be defect and must be changed For an accu change the unit must be sent...

Page 10: ...3 5 Charge accu connect the provided charging unit type ACS 110 with the connector of the ACS 1000 RC connect charging unit with the electrical network The red LED Charge indicates the charging proces...

Page 11: ...tall the Software net Framework from Microsoft is required Is it not yet installed it will be automatically downloaded and installed during the installation Internet connection is required On the CD a...

Page 12: ...disrupted 2 Transmit data record to table On the chosen measuring points each data record must be transmitted to the table 3 Print chart A window appears for customer and machine data Enter the custom...

Page 13: ...ner Huber eMail Info Hubitron de www Hubitron de 02 08 2012 Manufacturer Hubitron Steuerungstechnik Hauptstr 4 1 89611 Obermarchtal Germany Product Alignment Calibration System ACS 1000 and Remote Con...

Page 14: ...sanleitung ACS 1000 Rev 1 0 08 12 Hubitron Steuerungstechnik Hubitron Steuerungstechnik Hauptstr 4 1 D 89611 Obermarchtal Tel 49 0 7375 922066 FAX 49 0 7375 922077 Seite 14 von 14 Attachment Measuring...