Introduction
Chapter 1
OPERATION MANUAL
V2.3.0en/07.06.19//17.12
Unichiller®
19
1.4.3
Further protective devices
Emergency strategy – isolate the power supply!
To determine the type of switch or switch combination your temperature control unit is equipped
with, please refer to the connection sketch starting on page 85 in the section
.
Temperature control units with >Mains isolator<
[36]
(red/yellow or gray):
Turn the
>Mains
isolator<
[36] to the “0” position!
Temperature control units with >Mains isolator
[36]
(red/yellow) and additional >Appliance
switch<
[37]
(gray):
Turn the
>Mains isolator<
[36] to the “0” position! Then turn the
>Appliance
switch<
[37] to the “0” position!
Temperature control units with >Mains isolator<
[36]
(gray) and >Emergency stop switch<
[70]
(red/yellow):
Press the
>Emergency stop switch<
[70]. Then turn the
>Mains isolator<
[36] to the
“0” position!
Temperature control units with >Mains switch<
[37]
:
Power supply via socket: Disconnect the
temperature control unit from the power supply. Then turn the
>Mains isolator<
[37] to the “0”
position! Power supply via hard wiring: Disconnect the power supply by means of the building’s
circuit breaker. Then turn the
>Mains isolator<
[37] to the “0” position!
Temperature control units without a switch or inside a protective housing:
Connection via sock-
et: Disconnect the temperature control unit from the power supply. Connection via hard wiring:
Disconnect the power supply by means of the building’s circuit breaker.
1.4.3.1
Power interruption
Following a power outage (or when switching on the temperature control unit), this function can be
used to determine how the temperature control unit is supposed to respond. This response can be
determined via “Pilot ONE”.
OFF/Standby
(Default setting)
After turning the temperature control unit on, thermoregulation is started only after manual input.
ON / Temp. control active
After turning on the temperature control unit, thermoregulation is always started. An INFO appears
for a few seconds. This provides for suppressing the automatic start.
Only valid for temperature control units with an emergency stop switch (optional):
After installing the
>Emergency stop switch<
[70], automatic starting of the temperature control
process is
not possible
when turning on the temperature control unit.
Power Failure Auto
If temperature control is active during a power outage, the process will automatically continue after
the power outage.
Only valid for temperature control units with an emergency stop switch (optional):
After installing the
>Emergency stop switch<
[70], automatic continuation of the temperature con-
trol process is
not possible
after a power outage.
1.4.3.2
Alarm functions
An alarm is a system state that signals unfavorable process conditions. The temperature control unit
can be programmed so that the plant operator is warned when defined limit values are exceeded.
The response of the temperature control unit to an alarm can be determined. Possible responses
are: Switch off temperature control or control temperature to a safe setpoint (2nd setpoint).
1.4.3.3
Warning messages
Warning messages contain a message about the irregularity of the temperature control unit. These
messages have no further consequences. The plant operator evaluates the relevance of the message
and takes action where necessary.
1.4.3.4
Emergency stop switch (optional)
The
>Emergency stop switch<
[70]
immediately
disconnects all poles of the temperature control
unit. Further information on the
>Emergency stop switch<
[70] can be found on page 46 in Section
»Emergency stop switch (optional): Activate / Deactivate«
.