Setup mode
Chapter 4
OPERATION MANUAL
V2.3.0en/07.06.19//17.12
Unichiller®
47
The following data entry is required only for:
a) initial startup
b) reseting the temperature control unit to the factory settings
(see page 43, section
).
After switching on the temperature control unit, tap on the required system language.
Confirm your choice by tapping on “OK”.
Tap on the thermal fluid used.
Confirm your choice by tapping on “OK”.
Read the message and confirm by tapping on “OK”.
Enter the volume using the number keypad that appears.
Confirm your entry by tapping on “OK”.
If your temperature control unit is not integrated into a network, confirm the preset IP address
(0.0.0.0) by tapping on “OK”. This skips the network settings.
Enter the required IP address using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Enter the required subnet mask using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Tap on the required remote control mode.
Confirm your choice by tapping on “OK”.
Switch the temperature control unit on and off using the
>Mains isolator<
[36].
Only valid for temperature control units with an emergency stop switch (optional):
The safety relay is tested. Press the illuminated
>Release button<
[0]. The temperature control
unit is started and the illumination in the
>Release button<
[0] is turned off.
Only valid for temperature control units with an emergency stop switch (optional):
Proceed as follows, if the
>Release button<
[0] is not illuminated when turning on the tempera-
ture control unit via the
>Mains switch<
[36]:
Check whether the temperature control unit is properly
connected to a functioning
power supply.
Solution: Properly connect the temperature control unit to a working power supply.
Check whether the
>Emergency stop switch<
[70] has been activated (on page 46 see Section
»Emergency stop switch (optional): Activate / Deactivate«
). Solution: Deactivate the
>Emergency
stop switch<
[70] as shown on page 46 in Section
»Emergency stop switch (optional): Deacti-
.
If the illuminated
>Release button<
[0] does not turn off when pressed and/or the temperature
control unit still can’t be turned on:
Please contact our Customer Support (phone number, see page 84 in Section
).
4.1.3
Turning off the temperature control unit
PROCEDURE
Warm the thermofluid to room temperature.
Stop the thermoregulation.
Switch off the temperature control unit using the
>Mains switch<
[36].