3
III. technical parameters of series products
Ground
voltage generator
Model
Rated
Voltage
Load
Fuse
Weight
40kV
0.1Hz,≤1.1µF
Controller
:
6Kg
40KV /1.1
8A
0.05Hz,≤2.2µF
(
Peak
Value
)
Booster
:
20Kg
0.02Hz,≤5.5µF
50kV
0.1Hz,≤1.1µF
Controller
:
6Kg
50kV/1.1
10A
0.05Hz,≤2.2µF
(
Peak
Value
)
Booster
:
45Kg
0.02Hz,≤5.5µF
80kV
0.1Hz,≤1.1µF
80KV/0.5
20A
Controller
:
4Kg
0.05Hz,≤2.2µF
(
Peak
Value
)
booster
:
50Kg
0.02Hz,≤5.5µF
Compare content
Power frequency
withstand voltage
Frequency
withstand
voltage
0.1Hz withstand
voltage
DC withstand
voltage
Equivalence
good
Good
Good
Bad
Insulation
destructive
small
small
small
big
Operation safety
low
low
high
A little low
Wiring
Complicated
Most
Complicated
Most easy
complicated
Volume and
weight
Heavy and big
Heavy and big
Light and small
size
Light and small
size