4

“(OFF) Electric potential /Contact” use for Electric potential off or Contact off.

2.LCD can display the measured output voltage and current, action time and contact or

potential state;

Line 1: Select Grade display output voltage and current value;

Line 2: Stopwatch measured value Select Grade display output voltage and current value;

Stopwatch measure value display on the left, current status of stopwatch input terminal

display on the right;

Line 3: Left:Display external start or internal start, Right:

And the table state;

Line 4: Left:Display the current states are single or dual , Right: The stopwatch

state.

3.Resistors and Capacitors:

1

Resistance: 0.5Ω·10A

1Ω·5A

25Ω·2A

100Ω·0.7A

500Ω·0.3A

1000Ω·0.22A

2500Ω·0.14A; Obtain the resistance by select different terminal; accurately adjust the

required value of voltage and current by string into the voltage and current output grade;

2

Capacitor: 10uF/AC450V.

4.Measure selection indicate: 0~100A

0~40A

0~10A

0~250V

0~300V

external input;

Led Indicator Lights Up shows output value in LCD.

5.Split/join the auxiliary contact; the contact synchronize with 10 point "ON/ OFF" button.

6.AC voltage and current output terminal: 0~10A·90V

0~40A·25V

0~100A·10V

0~250V·3A

The black terminal is COM.

7.DC current output terminal: 0~300V·3A,the terminal is independent output Red + Black

-.

8.External input VOLTMETER & AMMETER; display measure value by select external

input grade of “Output selection”button.

9.Current and voltage output Knob: Adjust output value by match “ON/ OFF” of 10 contact

button.

10.“Output Selection” and “ON/ OFF” button: “Output Selection” button is only used in the

removal state match measure selection indication; “ON/ OFF” button is used for

controlling output and removal, Top left LED light up when output, lights out when

removal.

11.Auxiliary DC voltage power: can finely adjust voltage value independently, Select grade

Summary of Contents for HZJB-D

Page 1: ...HZJB D Single Phase Relay Protection Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...nual and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this in...

Page 3: ...Contents I General 1 II Main Features 1 III Main Technical Parameter And Functions 1 IV Panel And Function Instruction 3 V Usage Method 5 VI Caution 7 VII Packing List 7...

Page 4: ...es 1 Output a full isolation way for adjustable DC voltage Automatic protection for short circuit overload and out range 2 Maximum Output Current 0 150A 3 Equipped with LCD digital AC DC volt ampere m...

Page 5: ...0 000 6 000A AC or DC Accuracy 0 5 Measurement methods AC True RMS value DC Average 6 Voltmeter Range 0 0 600 0V AC OR DC Accuracy 0 5 Measurement methods AC True RMS value DC Average 7 Resistors and...

Page 6: ...tial Contact Reset for time reset check if the stopwatch reset to zero before counting each time Single Dual Channel for choosing stopwatch status On default external dual channel Press and hold more...

Page 7: ...ng into the voltage and current output grade 2 Capacitor 10uF AC450V 4 Measure selection indicate 0 100A 0 40A 0 10A 0 250V 0 300V external input Led Indicator Lights Up shows output value in LCD 5 Sp...

Page 8: ...voltage and current output 2 Auxiliary Output DC voltage and current First cycles the potentiometer to zero connect the output line select the required voltage value 0V or 110 220V then press output b...

Page 9: ...ot charged or charged contact switch signal Figures 1 1 4 are shown Note Regardless of the figure 1 3 4 connection way the black terminal of stopwatch should always connect to LOW level 3 Application...

Page 10: ...eep safety 2 Please adjust the contrast adjustment if the display shows dark or black 3 The measuring selection button is invalid after the current outputting VII Packing List No Item Qty 1 Main engin...

Reviews: