background image

2

0~300(DC)

320

250

3

3.DC auxiliary Output

Range V

Voltage V

Maximum current A

0~110

120

0.5

110~220

220

0.5

4.Stopwatch:

Range: 0.0000~999999S,Resolution: 0.1mS,

Accuracy: ±5

With electrical contacts and potential input

,

the maximum input voltage: DC 250V

5.Ammeter:

Range: internal: 0.000~150.0A;

External: 0.000~6.000A(AC or DC)

Accuracy: 0.5%, Measurement methods: AC: True RMS value DC: Average

6.Voltmeter:

Range: 0.0~600.0V(AC OR DC);

Accuracy: 0.5%, Measurement methods: AC: True RMS value DC: Average

7.Resistors and Capacitors:

Resistance: 0.5 ohm to 2.5k ohm

Capacitor: 10uF, maximum voltage AC 450V

8.Division / combination of auxiliary contacts

Maximum current: 1A, Maximum voltage: AC250V or DC120V

9.Using Environment: Temperature: -10 ~ 40

,

Relative humidity must less than 80%

10.Volume: 320mm×260mm×230mm

11. Weight: 16kg

Summary of Contents for HZJB-D

Page 1: ...HZJB D Single Phase Relay Protection Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...nual and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this in...

Page 3: ...Contents I General 1 II Main Features 1 III Main Technical Parameter And Functions 1 IV Panel And Function Instruction 3 V Usage Method 5 VI Caution 7 VII Packing List 7...

Page 4: ...es 1 Output a full isolation way for adjustable DC voltage Automatic protection for short circuit overload and out range 2 Maximum Output Current 0 150A 3 Equipped with LCD digital AC DC volt ampere m...

Page 5: ...0 000 6 000A AC or DC Accuracy 0 5 Measurement methods AC True RMS value DC Average 6 Voltmeter Range 0 0 600 0V AC OR DC Accuracy 0 5 Measurement methods AC True RMS value DC Average 7 Resistors and...

Page 6: ...tial Contact Reset for time reset check if the stopwatch reset to zero before counting each time Single Dual Channel for choosing stopwatch status On default external dual channel Press and hold more...

Page 7: ...ng into the voltage and current output grade 2 Capacitor 10uF AC450V 4 Measure selection indicate 0 100A 0 40A 0 10A 0 250V 0 300V external input Led Indicator Lights Up shows output value in LCD 5 Sp...

Page 8: ...voltage and current output 2 Auxiliary Output DC voltage and current First cycles the potentiometer to zero connect the output line select the required voltage value 0V or 110 220V then press output b...

Page 9: ...ot charged or charged contact switch signal Figures 1 1 4 are shown Note Regardless of the figure 1 3 4 connection way the black terminal of stopwatch should always connect to LOW level 3 Application...

Page 10: ...eep safety 2 Please adjust the contrast adjustment if the display shows dark or black 3 The measuring selection button is invalid after the current outputting VII Packing List No Item Qty 1 Main engin...