background image

3

2.10 Wiring mode: as required

2.11 Protection settings: overcurrent, overvoltage

III.Working principle

1. Functional block diagram

2. Working principle

The instrument measurement circuit includes a current measurement circuit and a current

measurement circuit. (Voltage test can be added as required) The volt-ampere

characteristic function can be expanded.

The current measurement loop includes micro-current zero-impedance CT, programmable

amplifying circuit and sampling circuit.

The voltage measurement circuit includes a PT isolation signal acquisition circuit, a

programmable amplifier circuit and a sampling circuit.

The 16-bit single-chip microcomputer uses computer digital real-time acquisition method,

by measuring the voltage signal amplitude, according to the voltage proportional

relationship, the high-voltage measured voltage value can be calculated, and the

automatic timing function can be realized by comparing the measured voltage signal

amplitude with the set voltage value. According to the current proportional relationship, the

Reviews: