background image

4

SWERV

® 

INSTALLATION AND ASSEMBLY

HANDPIECE CORD CONNECTION

Align the marks on the control unit and the handpiece cord 

plug. Firmly push the plug into the connector at the front of the 

control unit. (Fig. 1)

CAUTION: 

1.   Do not disconnect the handpiece cord when power to device 

is switched on and the footswitch is depressed.

2.   Detach insert before removing handpiece cord.
3.   Grip the plug of the handpiece cord when removing.

ULTRASONIC INSERT CONNECTION

Select appropriate kHz insert (25 or 30) for the generator. 

Fill handpiece with water. While holding the handpiece in 

an upright position, activate the purge function or the foot 

control to remove any air bubbles that may be trapped inside 

the handpiece. Lubricate the rubber o-ring on the sterilized 

insert with water prior to placing it into the handpiece. Fully 

seat the insert with a gentle push-twist motion. Do not force 

the insert into the handpiece. 

WARNING:

 Always use water when operating the unit to keep 

the handpiece and insert cool. Minimum recommended setting 

>10 m l/min. Maximum recommended setting < 50 ml/min. 

FOR OPERATION

  Verify that the water supply is adequate and clean prior to 

beginning each procedure. 

 Purge handpiece.

  Use a Hu-Friedy disposable barrier sleeve (UMBS) on the handpiece. Replace the barrier between 

patients. Properly dispose of the barrier in accordance with guidelines for contaminated waste 

materials. 

 Place insert in handpiece.

 Adjust the power level to the recommended mode. 

 Depress the foot control to start insert vibration. 

  Turn the water adjustment knob counterclockwise gradually to increase the water flow. (Fig. 2) 

WHEN TREATMENT IS COMPLETED

1.   Release the foot control and place the handpiece in the handpiece holder. (Fig. 3)
2.  Turn off the power switch.
3.   Hold handpiece in upright position. Gently pull the insert out of the handpiece opening.  

Ultrasonic inserts are easily interchangeable for various applications and procedures.

4. Close the dental unit’s water valve. 

5

1

7

2

8

3

Summary of Contents for SWERV3

Page 1: ...MAGNETOSTRICTIVE SCALER HANDPIECE User Manual Manuel de l utilisateur SWERV 3...

Page 2: ......

Page 3: ...al treatment of Periodontal diseases Read this operation manual carefully before use and retain for reference SYMBOLS AND GENERAL SAFETY Attention consult accompanying documents Risk of damage or harm...

Page 4: ...ert into the handpiece WARNING Always use water when operating the unit to keep the handpiece and insert cool Minimum recommended setting 10 m l min Maximum recommended setting 50 ml min FOR OPERATION...

Page 5: ...ilize the handpiece or handpiece cord PROCEDURES FOR HU FRIEDY ULTRASONIC INSERTS sold separately 1 Clean inspect and steam sterilize before each use 2 For complete instructions refer to cleaning and...

Page 6: ...ases of the Handpiece having been used outside of normal operating conditions the Handpiece being damaged in transit the Handpiece having been modified non compliance with the manufacturer s installat...

Page 7: ...tales Lire ce manuel d exploitation soigneusement avant usage et le conserver pour r f rence SYMBOLES ET S CURIT G N RALE Attention consulter les documents joints Risque d endommagement ou de blessure...

Page 8: ...t dans la pi ce main AVERTISSEMENT toujours utiliser de l eau pour refroidir la pi ce main et l insert lorsque l appareil est en marche R glage minimal recommand 10 ml min R glage maximal recommand 50...

Page 9: ...RDE ne pas st riliser l unit de commande le cordon d alimentation la commande au pied la conduite d eau le connecteur d irrigation la pi ce main ou le cordon de la pi ce main PROC DURES POUR LES INSER...

Page 10: ...la pi ce main en dehors des conditions normales de fonctionnement endommagement de la pi ce main en transit modification de la pi ce main non conformit aux instructions d installation du fabricant pr...

Page 11: ...11...

Page 12: ...Hu Friedy Mfg Co LLC 3232 N Rockwell St Chicago IL 60618 USA Telephone 1 800 Hu Friedy or 1 773 975 6100 Website Hu Friedy com 2018 Hu Friedy Mfg Co LLC All rights reserved UMMANUALHP 0818...