THT60 - THT70
EN - 26
4. Touch the field “
Reset parameters
” to restore the default conditions on the instrument.
The following screen appears on the display:
Fig. 38: Restoring default conditions on the instrument
5. Touch the “Yes” key to restore the default conditions (see Table 1) or the “No” key to
cancel the operation.
Parameter Default
setting
Emissivity
0.95
Distance 5m
Environmental temperature
25°C
Relative humidity
60%RH
Reflected temperature
25°C
Screening Active
Table 1: Thermal camera default settings
Setting the general measuring parameters
1. Touch the item “
Measure Setting
”. The following screen appears on the instrument's
display:
Fig. 39: Setting the general measuring parameters
2. The parameters whose field is included within the symbols “
<
” and “
>
” can be set
¾
Lens
Æ
Indication of the type of IR lens installed on the instrument. This parameter
is automatically modified by the instrument by changing the lens (see § 4.2.7).
Possible options:
7.5mm
(default),
11mm
,
22mm
,
33mm
(THT60),
22mm
(default),
11mm
,
38mm
(THT70)
¾
Range
Æ
Indication of the temperature measuring range. Options:
-20°C
y
150°C
and
0
y
400°C
¾
Env
:
Æ
Environmental temperature setting Measuring range:
-40°C
y
125°C
¾
Humis
Æ
Relative humidity setting. Measuring range:
1%RH
y
100%RH
¾
Ref
Æ
Reflected temperature setting (see § 5.2.2.5). Range:
-100°C
y
2000°C
3. Press the “
” key to go back to the previous screen