5.7. POSITION
“
Ω
”: RESISTANCE AND CONTINUITY MEASUREMENT
WARNING
Before attempting any resistance measurement, remove power from the
circuit under test and discharge all capacitors, if present.
By positioning the selector switch to “
Ω
”, the screen to the side
will appear.
Mod OK
Fnz
>
60.0 k
Ω
19/01-17:00:00
Fig. 13: resistance measurement
Insert the red cable into the input lead
V
Ω
and the black cable into the input lead
COM
,
and connect the instrument as described in Fig. 13.
Press key
F1
(
Mod
) to open the drop-down menu shown on the
screen to the side. At each subsequent pressure of key
F1
, the
cursor will scroll through the available items, as follows:
•
Resistance
: resistance measurement;
•
Continuity
: continuity measurement;
•
Help
: it displays the connection between instrument and
system;
•
Esc
: it closes the drop-down menu.
Mod
OK
Fnz
>
60.0 k
Ω
19/01-17:00:00
Select the desired mode and press key
F3
(
OK
) to confirm.
Resistance
Continuity
Help
Esc