5.3. POSITION
“V ”: DC, AC+DC VOLTAGE MEASUREMENT AND PHASE
SEQUENCE VERIFICATION
WARNING
The maximum DC or AC+DC input voltage is 1000V. When the display
shows “> 999.9V”, it means that the maximum value the clamp is able to
measure has been exceeded. Exceeding these limits could result in
electrical shocks to the user and damage to the instrument.
By positioning the selector switch to “
V
”, the screen to the side
will appear.
Mod Par Fnz
A C < 1 0 . 0 H z
- - - -
V
19/01-17:00:00
Press key
F1
(
Mod
) to open the drop-down menu shown on the
screen to the side. At each subsequent pressure of key
F1
, the
cursor will scroll through the available items, as follows:
•
AC
: AC+DC voltage measurement;
•
DC
: DC voltage measurement;
•
Ph Seq
: verification of phase sequence;
•
Help
: it displays the connection diagram of the instrument to
the system;
Mod
Par OK
A C < 1 0 . 0 H z
- - - -
V
19/01-17:00:00
Select the desired mode and press key
F4
(
OK
) to confirm.
5.3.1. DC voltage measurement
Fig. 5: DC voltage measurement
AC
DC
Ph Seq
Help
Esc