EQUITEST
EN - 8
5 OPERATING INSTRUCTIONS
5.1 CONTINUITY MEASUREMENT OF PROTECTIVE CONDUCTORS WITH 10A
CAUTION
The power supply voltage of the accessory must be: 230/240V ± 10% (207V
÷ 265V) or 110V ± 10% (100V ÷ 120V). Do not use external power supply
voltages external from the above ranges to avoid damaging the instrument
Check that the rated operating voltage value indicated on the back
label should be consistent with the power supply voltage of the
system under test
Fig. 3: Continuity measurement of protective conductors with 10A
1. Connect the accessory to the power supply using the supplied network cable (see Fig.
1 - part 10). The
POWER
LED is
steady green
to identify the correct power supply
voltage, the
STATUS
LED is
steady green
to identify the operation of the accessory in
Manual mode (see user manual of the MASTER instrument) and the
WiFi
LED is on in
green flashing
color to identify that the accessory is trying to make (if available) a WiFi
connection with an external mobile device via the HTAnalysis APP (refer to the
HTAnalysis instructions for managing the operation)
2. Connect the accessory to the MASTER instrument (see Fig. 3) by using the C2050
cable supplied (see Fig. 1 - part 1)
3. Check the correct detection of the accessory on the MASTER instrument (see relevant
user manual)
Summary of Contents for EQUITEST
Page 3: ... Copyright HT ITALIA 2022 Versione IT 1 00 19 04 2022 ITALIANO Manuale d uso ...
Page 17: ... Copyright HT ITALIA 2022 Release EN 1 00 19 04 2022 ENGLISH User manual ...
Page 31: ... Copyright HT ITALIA 2022 Versión ES 1 00 19 04 2022 ESPAÑOL Manual de instrucciones ...
Page 45: ... Copyright HT ITALIA 2022 Ausführung DE 1 00 19 04 2022 DEUTSCH Bedienungsanleitung ...
Page 59: ... Copyright HT ITALIA 2022 Version FR 1 00 19 04 2022 FRANÇAIS Manuel d utilisation ...
Page 73: ... Copyright HT ITALIA 2022 Versão PT 1 00 19 04 2022 PORTUGUÊS Manual de instruções ...
Page 87: ......