background image

 

EN

G

 

 

HT12 

 

USER MANUAL

 

1. PRECAUTIONS AND SAFETY MEASUREMENTS

 

This  instrument  complies  with  safety  Standard  IEC/EN61010-1 
related to electronic measuring instruments. For your own safety and 
to avoid damaging the instrument follow the procedures described in 
this  instruction manual and read carefully all notes preceded by this 

symbol 

 

 

 

CAUTION 

  Avoid measuring in humid or wet places 
  Avoid  measuring  in  rooms  where  explosive 

gas, combustible gas, steam or excessive dust 
is present 

  Keep you insulated from the object under test 
  Do not touch exposed metal parts such as test 

lead ends, sockets, fixing objects, circuits etc. 

  Avoid  doing  that  if  you  notice  anomalous 

conditions  such  as  breakages,  deformations, 
fractures, leakages of battery fluid, blind display 
etc. 

  Be careful when measuring voltages exceeding 

20V to avoid risks of electrical shock 

 

The following symbols are used in user manual and on the meter: 

 

 

CAUTION  -  refer  to  the  instruction  manual  -  an 
improper  use  may  damage  the  instrument  or  its 
components 

 

 

 

Double insulated meter

 

 

 

 

The  instrument  can  clamp  not  insulated  live  cables 
corresponding 

to 

the  measurement 

category 

indicated on them 

 

 

 

AC voltage or current 

 

 

 

DC voltage or current 

 

 

 

CAUTION

:  this  symbol  indicates  that  equipment,  its 

accessories  and  battery  shall  be  subject  to  a 
separate collection and correct disposal 

 

 

1.1 Preliminary instructions 

 

CAUTION 

  This  instrument  has  been  designed  for  use  in 

environments with pollution degree 2 

  It  can  be  used  for 

VOLTAGE

  and 

CURRENT

 

measurements

 

on installations with overvoltage 

category CAT III 300V and CAT II 600V 

  Only  the  leads  supplied  with  the  instrument 

guarantee 

compliance 

with 

the 

safety 

standards  in  force.  They  must  be  under  good 
conditions  and,  if  necessary,  replaced  with 
identical ones 

  Do not test or connect to any circuit exceeding 

the specified overload protection 

  Do 

not 

effect 

measurements 

under 

environmental  conditions  exceeding  the  limits 
indicated in § 8 

  Make sure that batteries are properly installed 
  Before  connecting  the  test  probes  to  the 

installation,  make  sure  that  the  function 
selector 

is 

positioned 

on 

the  required 

measurement 

  Make  sure  that  LCD  and  function  selector 

indicate the same function 

  When  changing  the  range,  first  disconnect  the 

test leads from the circuit under test in order to 
avoid any accident 

  When 

the 

instrument 

is 

connected 

to 

measuring  circuits  never  touch  any  unused 
terminal 

  If  during  measurement  the  displayed  values 

remain  constant  check  whether  the  HOLD 
function is active 

 

1.2 Measuring overvoltage category definitions 

According to the IEC/EN61010-1 guidelines the circuits are divided 
into the following measurement categories: 

 

CAT IV

 is for measurements performed at the source of the low-

voltage installation 

 

CAT

 

III

 is for measurements performed in the building installation 

 

CAT  II

  is  for  measurements  performed  on  circuits  directly 

connected to the low voltage installation 

 

CAT  I

  is  for  measurements  performed  on  circuits  not  directly 

connected to MAINS 

 

2. GENERAL DESCRIPTION

 

HT12 performs the following measurements: 

 

 DC and AC voltage 
 DC and AC current by using integrated sensor fork clamp 
 Resistance and continuity test 
 Voltage and current frequency 

 

All selectable by means of a 7-position function selector. Function 
keys 

DH 

and 

RH 

in  addition  to  an  analogical  bargraph  are  also 

available (see § 5.1 and § 5.2). The selected quantity is displayed 
with  indication  of  measuring  unit  and  active  functions.  The 
instrument is provided with an Auto Power Off function consisting 
in  an  automatic  switching  off  10  minutes  after  last  pressure  on 
keys  or  rotation  of  selector. To resume  normal operation,  switch 
on the instrument again 

 

3. PREPARATION FOR USE 

3.1 Initial checks

 

This  instrument  was  checked  both  mechanically  and  electrically 
prior to shipment. All possible cares and precautions were taken 
to  let  you  receive  the  instrument  under  perfect  conditions. 
Notwithstanding we suggest you to check it rapidly (any damage 
may  have  occurred  during  transport 

–  if  so  please  contact  the 

local distributor from whom you bought the item). Make sure that 
all  standard  accessories  mentioned  in  §  8  are  included.  Should 
you  have  to  return  back  the  instrument  for  any  reason  please 
follow the instructions mentioned in § 9 

 

3.2 Supply voltage

 

The  instrument  is  powered  by  2x1.5V  batteries  AAA  LR03 
included  in  the  packaging.  When  batteries  are  low  the  symbol 
"

"  is  displayed.  To  replace/insert  batteries  follow  the 

instructions indicated in § 7 

 

3.3 Calibration

 

The  instrument  complies  with  the  technical  specifications 
contained in this manual and such compliance is guaranteed for 
12 months 

 

3.4 Storage 

After a period of storage under extreme environmental conditions 
exceeding  the limits  mentioned  in  §  8 let  the  instrument  resume 
normal operating conditions before using it 

4. INSTRUMENT AND DISPLAY DESCRIPTION 

 

 

 

 

 

                         

                        

                   

 

 Copyright HT ITALIA 2013                                  Release EN 1.02 - 15/02/2013 

5. FUNCTION KEYS DESCRIPTION 

5.1 DH key

 

By pressing 

DH

 key the measured value is hold on the display where 

the symbols "H" 

and “ ” appear. Pressing again 

DH

 to disable this 

function and resume normal operation 

 

5.2 RH key

 

By  pushing 

RH

  key,  t

he  manual  mode  is  activated,  the  “AUTO” 

symbol  disappears 

and  the  symbol  “ ”  is  displayed.  To  quit  this 

function  keep 

RH

  key  pressed  for  at  least  1  second  or  rotate  the 

selector  into another position. The Autorange mode is  automatically 
active whenever you switch on the meter 

 

6. HOW TO PERFORM THE MEASUREMENTS

 

6.1 DC/AC Voltage measurement 

 

CAUTION 

Maximum  input  for  AC/DC  voltage  measurements  is 
600V.  Do  not  take  any  voltage  measurement 
exceeding this limit in order not to risk electrical shock 
or damaging the meter 

 

1.  Rotate  the  selector  on 

  or 

position. 

The  “DC”  or  “AC”  and 

“AUTO”  symbols  are  shown  at  display.  Press 

RH

  key  for  the 

selection of manual range (see § 5.2) 

2.  Connect the red and black test leads to the positive and negative 

poles  of  the circuit  under  test  respectively  (for DC  voltage). The 
voltage value is displayed 

3.  If  the  message  "

O

.

L

"  is  displayed  the  maximum  voltage  value 

was exceeded 

4.  On  DC  voltage  measurement  the  symbol  "

-

"  on  the  instrument 

display indicates that voltage has opposite direction compared to 
the previous connections 

5.  For Data HOLD feature see § 5.1 

 

6.2 Current measurement with fork sensor clamp 

 

CAUTION 

 Do  not  perform  measurements  on  circuits  with  a 

voltage exceeding 600V AC/DC between phases or 
exceeding 300V AC/DC between phase and ground 
to avoid possible shock hazard 

 Do  not  perform  measurements  with  test  leads 

connected to the circuit under test 

 Do  not  apply  strong  vibrations  to  the  clamp  sensor 

and  keep  the  fingers  behind  the  protective  bar 
during measurements 

 

 

 

6.2.1 DC Current measurement 

1.  Rotate the selector on 

position

The “DC” and “AUTO” symbols 

are  displayed.  Press 

RH

  key  for  the  selection  of  manual  range 

(see § 5.2) 

2. 

Perform  zeroing  operation  turning  the  adjust  knob

  which  is 

present on the central part of the  clamp sensor. The 

“0.0” value 

should be displayed to perform correct measurements 

3.  Put the conductor in the fork of clamp sensor at the centre of the 

arrows  keys  which  are  present  in  the  frontal  side  (see  picture) 
and  respect  the  current  direction  as  indicated  by  the  arrow  key 
which is present on the meter sideways 

4.  The current value is displayed. The symbol "

-

" on the instrument 

display  indicates  that  the  current  has  opposite  direction 
compared to the previous connection 

6.2.2 AC Current measurement 

1.  Rotate  the  selector  onto 

position

The  “AC”  and  “AUTO” 

symbols are displayed. Press 

RH

 key for the selection of manual 

range (see § 5.2) 

2.  Put the conductor in the fork of clamp sensor at the centre of the 

arrows  keys  which  are  present  in  the  frontal  side  (see  picture) 
and  respect  the  current  direction  as  indicated  by  the  arrow  key 
which is present on the meter sideways 

3.  The current value is displayed 
4.  For Data HOLD feature see § 5.1 

6.3 Resistance and Continuity test measurements 

 

CAUTION 

Before  taking  resistance  measurements  on  circuit 
remove  power  from  the  circuit  under  test  and 
discharge all capacitors 

1.  Rotate  the  selector  onto 

  / 

  position. 

The  “M ”  and  “

O.L

” 

symbols are displayed. Press 

RH

 key for the selection of manual 

range (see § 5.2) 

2.  Connect  the  test  leads  to  the  circuit  under  test.  The  resistance 

value is displayed 

3.  The continuity test is always active and the buzzer sounds when 

resistance  values  are  lower  than  30 .  In  Manual  mode  the 

“ ” 

symbol is displayed only in the 340  range 

 

6.4 Frequency measurement

 

 

CAUTION 

Do  not  perform  measurements  on  circuits  with  a 
voltage  exceeding  600V  AC/DC  between  phases  or 
exceeding  300V  AC/DC  between  phase  and  ground 
to avoid possible shock hazard 

6.4.1 Voltage frequency measurement

 

1.  Rotate  the  selector  on 

Hz

  position.  Press 

RH

  key  for  the 

selection of manual range (see § 5.2) 

2.  Connect the test leads to the circuit under test. The frequency 

value (in kHz or MHz) is displayed 

6.4.2 Current frequency measurement

 

1.  Rotate  the  selector  on 

Hz

  position.  Press 

RH

  key  for  the 

selection of manual range (see § 5.2) 

2.  Put the conductor in the fork of clamp sensor at the centre of 

the arrows keys. The value (in kHz or MHz) is displayed 

 

DO NOT CONNECT SIMULTANEOUSLY THE TEST LEADS 

AND THE CLAMP SENSOR TO THE INSTALLATION 

 

 

7. BATTERY REPLACEMENT

 

When  the  "

"  low  battery  indication  is  displayed  the  batteries 

must be replaced as follows: 

 

1.  Switch 

OFF

 the meter 

2.  Remove the protection holster from the instrument 
3.  Remove the fixing screw and the battery compartment cover 
4.  Remove batteries and replace with new ones of the same type 
5.  Replace battery compartment cover and holster 

 

 

 

8. TECHNICAL SPECIFICATIONS 

The accuracy is referred to 23°C± 5°C, <75%RH 

AC Current 

Overload protection: 72A AC/DC for 10s 

DC Current 

Overload protection: 72A AC/DC for 10s

 

AC Voltage 

Input impedance: 10M  ; Overload protection: 720V AC/DC for 10s 

DC Voltage 

Input impedance: 10M  ; Overload protection: 720V AC/DC for 10s 

Resistance and Continuity test 

Frequency 

Current sensitivity: 15A ; Voltage sensitivity: 30V 

 

Safety: IEC/EN61010-1, IEC/EN61010-2-031, IEC/EN61010-2-032 
Measurement category: CAT III 300V , CAT II 600V, Pollution degree: 2 
Insulation: double insulation 
Compliance with 2004/108/CE (EMC) and 2006/95/CE (LVD) Directives 
Display: LCD, ¾ dgt, 3400 points with bargraph 

Meas. Range 

Resolution 

Accuracy 

0.0   60.0A 

0.1A 

(2.0%rd+5dgt) (50/60Hz) 

Meas. Range 

Resolution 

Accuracy 

0.0   60.0A 

0.1A 

(2.0%rdg+5dgt) 

Meas. Range 

Resolution 

Accuracy 

0.000   3.400V 

0.001V 

(1.5%rdg+5dgt) (50-400Hz) 

0.00   34.00V 

0.01V 

0.0   340.0V 

0.1V 

0   600V 

1V 

Meas. Range 

Resolution 

Accuracy 

0.000   340.0mV 

0.001mV 

(1.5%rdg+4dgt) 

0.000   3.400V 

0.001V 

0.00   34.00V 

0.01V 

0.0   340.0V 

0.1V 

0   600V 

1V 

Meas. Range 

Resolution 

Accuracy 

0.0   340.0  

0.1  

(1.0%rdg+3dgt) 

Buzzer active for R<30

10  

Only on 340  range 

0.000   3.400k  

0.001k  

0.00   34.00k  

0.01k  

0.0   340.0k  

0.1k  

0.000   3.400M  

0.001M  

(5.0%rdg+5dgt) 

0.00   34.00M  

0.01M  

(15.0%rdg+5dgt) 

Meas. Type 

Meas. Range 

Accuracy 

Current 

0   10kHz 

(0.1%rdg+1dgt) 

Voltage 

0   300kHz 

Sampling rate: approx 400ms (approx 20ms bargraph) 
Over range indication: “

O.L

” or “

-O.L

” symbols at display 

Power supply: 2x1.5V alkaline batteries type AAA LR03 UM-4 
Auto Power OFF: after 10 minutes of idleness (not disable) 
Reference temperature/humidity: 23°C± 5°C (73°F ± 41°F) ; <75%RH 
Working temperature/humidity: 0   40°C (32°F   104°F)  ; <85%RH 
Storage temperature/humidity: -20   60°C (-4°F   140°F) ; <85%RH 
Max height of use: 2000m (6562ft) 
Max diameter conductor : 6mm (0.2in) 
Dimensions (LxWxH): 128x87x24mm (5x3x1in) 
Weight (with batteries): 210g (7ounces) 
Included accessories: batteries, user manual  

 

9. SERVICE

 

9.1 Warranty conditions 

This  instrument  is  guaranteed  for  one  year  against  material  or 
production defects, in accordance with our general sales conditions.  
 
During  the  warranty  period  the  manufacturer  reserves  the  right  to 
decide  either  to  repair  or  replace  the  product.  Should  you  need  for 
any  reason  to  return  back  the  instrument  for  repair  or  replacement 
take  prior  agreements  with  the  local  distributor  from  whom  you 
bought it.  
 
Do not forget to enclose a report describing the reasons for returning 
(detected  fault). Use  only  original  packaging.  Any  damage occurred 
in transit due to not original packaging will be charged anyhow to the 
customer 

 

9.2 Service 

Should the instrument not work properly, make sure that batteries are 
correctly installed and working before contacting your distributor, also 
check the test leads and replace them if necessary.  
 
Should  you  need  for  any  reason  to  return  back  the  instrument  for 
repair or replacement take prior agreements with the local distributor 
from whom you bought it.  
 
Do not forget to enclose a report describing the reasons for returning 
(detected fault).  Use  only  original  packaging. Any  damage  occurred 
in transit due to not original packaging will be charged anyhow to the 
customer. The manufacturer will not be responsible for any damage 
to persons or things. 

 

 

HT ITALIA SRL

 

Via della Boaria, 40 
48018 

– Faenza (RA) – ITALY 

Tel: +39-0546-621002 
Fax: +39-0546-621144 
Web: 

www.ht-instruments.com

 

Email: 

[email protected]

  

 

www.

GlobalTestSupply

.com

Quality HT Products Online at:

[email protected]

Summary of Contents for 12

Page 1: ...erra al fine di evitare shock elettrici e possibili danni allo strumento Non eseguire misure con puntali di misura collegati all impianto Non applicare forti vibrazioni al sensore a pinza durante la misura e mantenere le dita al di sotto della barriera di protezione del sensore 6 2 1 Misura di Corrente DC 1 Selezionare la posizione I simboli DC e AUTO sono presenti a display Premere il tasto RH pe...

Page 2: ...hase and ground to avoid possible shock hazard Do not perform measurements with test leads connected to the circuit under test Do not apply strong vibrations to the clamp sensor and keep the fingers behind the protective bar during measurements 6 2 1 DC Current measurement 1 Rotate the selector on position The DC and AUTO symbols are displayed Press RH key for the selection of manual range see 5 2...

Page 3: ...itar el riesgo de sufrir un choque eléctrico no realice nunca medida en circutos con más de 600V CA CC entre cables y 300V CA CC entre conductor y tierra No realice mediciones con los cables de prueba conectados en el circuito en prueba Cuando utilice el sensor tenga la precaución de que no reciba golpes ni vibraciones excesivas y no efectúe mediciones con la mano después de la barrera de segurida...

Page 4: ...00V Phase gegen Erde Versuchen Sie keine Spannung zu messen die höher ist Es besteht die Gefahr eines Stromschlages Führen Sie keine Messungen durch wenn die Messleitungen am Messkreis angeschlossen sind Vermeiden Sie starke Vibrationen bei der Gabelstromzange und halten Sie die Finger hinter dem Schutzring während der Messung 6 2 1 DC Strommessung 1 Stellen Sie den Drehschalter in die Position Da...

Reviews: