IMP57
EN - 16
If the instrument is
overheated the display
will show the message
on side.
Please wait until the
LED STATUS on IMP57
get back to Green light.
LOOP 05.06.02
Z= - - - - -
R=-----
X=-----
IkSTD=----A
V1-2=461V FRQ=50.0HZ
HIGH TEMP
P-P Z2
FUNC ZSTD ICAL RMT
Instrument overheated
If You get the message
on side after pressing
START button, please
contact HT Service.
LOOP 05.06.02
Z= - - - - -
R=-----
X=-----
IkSTD=----A
V1-2=461V FRQ=50.0HZ
Error 5
P-P Z2
FUNC ZSTD ICAL RMT
PREVIOUS RESULT CAN'T BE STORED INTO INSTRUMENT MEMORY
.
If the instrument
measure an impedance
higher than 1999m
will
show the message on
side. Disable the
Z2
High Resolution
measurement mode and
execute standard LOOP
measurement.
LOOP 05.06.02
The ">" symbol indicate
that the measure is higher
than 1999m
Z> 1999
m
R>1999m
X>1999m
IkSTD=----A
V1-2=461V FRQ=50.0HZ
P-P Z2
FUNC ZSTD ICAL RMT
You can save these results by pressing the
SAVE
button
twice
.