4.3. DESCRIPTION OF FUNCTION KEYS
4.3.1. ON/OFF key ( )
Pressing the key allows turning on the instrument. Press and hold the key for approx.
1s to turn off the instrument.
4.3.2. AutoHold/
Pressing the
AutoHold/
key allows activating/deactivating the built-in white-light torch
(see Fig. 1
–
part 1) and the symbol
“
”
appears on the display. Press and hold the
AutoHold/
key for approx. 1s to activate/deactivate the
Auto HOLD
function. The
symbol
“
A HOLD
”
appears on the display and the buzzer beeps twice in a short time. With
activated Auto HOLD function, the instrument automatically freezes the measured value
on the display when the input signal is steady (3 samplings with a maximum difference of
10% between the highest and the lowest value). The instrument automatically quits the
function after approx. 30s seconds.
4.3.3. V/
Ω
/ key
CAUTION
The instrument switches on in
Ω
mode
.
The
instrument
automatically
switches to VAC or VDC if the following
voltages are detected:
VAC > ca. 1.5V ; VDC > ca. 1.0V
The instrument
automatically
switches to
Ω
mode with the following
measured voltages:
VAC and VDC < ca. 0.6V
Press the
V/
Ω
/
key to manually select resistance measurement and
continuity test.
4.3.4. Detection of AC voltage without contact
The instrument allows detecting the presence of AC voltage without contact by using the
NCV sensor found on the upper part (see Fig. 1
–
part 2).
1. Switch on the instrument by pressing the
ON/OFF
key.
2. Bring the NCV sensor closer to the potential AC source.
3. If the red detector turns on (see Fig. 1
–
part 3), AC voltage is present.
4.3.5. Auto-Power-Off
function
In order to preserve internal batteries, the instrument switches off automatically
approximately 15 minutes after it was last used. The symbol
“
”
identifies the Auto power
off function, which cannot be deactivated.