HP4402.book : chapter2.fm 40 Tue Apr 22 15:20:23 1997
Power Meter Operation
Recorder Output
2-40
HP EPM-441A User’s Guide
Leveling a Source Output
The Recorder output can be used to level an output from a source using
external leveling. The following procedure explains how to do this:
1.
Select the channel to be represented on the Recorder Output by
pressing
,
,
,
.
2.
The highest power you are going to measure is used to determine
the value which you should set for the Recorder Output maximum
setting. For example, if you are measuring a power less than 1 mW
and greater than 100
µ
W, then set the recorder maximum value to
1 mW.
To set the maximum value press
,
,
,
and enter the appropriate
value.
3.
Press
and enter 0 W.
4.
Press
to “
On
”.
50 dBm (100 W)
40 dBm (10 W)
30 dBm (1 W)
20 dBm (100 mW)
10 dBm (10 mW)
0 dBm (1 mW)
-10 dBm (100
µ
W)
-20 dBm (10
µ
W)
-30 dBm (1
µ
W)
-40 dBm (100 nW)
-50 dBm (10 nW)
-60 dBm (1 nW)
System
Inputs
More
Recorder Output
Output A B
System
Inputs
More
Recorder Output
Max Power
Min Power
Output Off On
Summary of Contents for EPM E4418A
Page 20: ...HP4402 book HP4402LOF doc 6 Tue Apr 22 15 20 23 1997 Contents 6 HP EPM 441A User s Guide ...
Page 22: ...HP4402 book HP4402LOT doc 8 Tue Apr 22 15 20 23 1997 Contents 8 HP EPM 441A User s Guide ...
Page 23: ...HP4402 book chapter1 fm 1 Tue Apr 22 15 20 23 1997 1 Getting Started ...
Page 41: ...HP4402 book chapter2 fm 1 Tue Apr 22 15 20 23 1997 2 Power Meter Operation ...
Page 101: ...HP4402 book chapter3 fm 1 Tue Apr 22 15 20 23 1997 3 Menu Reference ...
Page 127: ...HP4402 book chapter4 fm 1 Tue Apr 22 15 20 23 1997 4 Error Messages ...
Page 137: ...HP4402 book chapt5_2 fm 1 Tue Apr 22 15 20 23 1997 5 Specifications ...
Page 158: ...HP4402 book HP4402IX doc 6 Tue Apr 22 15 20 23 1997 Index 6 HP EPM 441A User s Guide ...