HP4402.book : chapt5_2.fm 10 Tue Apr 22 15:20:23 1997
Specifications
Power Meter Supplemental Characteristics
5-10
HP EPM-441A User’s Guide
For HP ECP-series power sensors
In FAST mode, within the range -50 dBm to +17 dBm, for a 10 dB
decreasing power step, the settling time is 10 ms
1.
1
When a decreasing power step crosses the power sensor’s auto-range switch
point, add 25 ms. Refer to the relevant power sensor manual for further
information.
For HP ECP-series power sensors in normal and x2 speed modes, manual
filter, 10 dB decreasing power step:
Table 5-5: Settling Time
Number of
Averages
1
2
4
8
16
32
64
128
256
512
1024
Settling Time (s)
(Normal Mode)
0.07
0.12
0.21
0.4
1
1.8
3.3
6.5
13
27
57
Response Time (s)
(x2 mode)
0.04
0.07
0.12
0.21
0.4
1
1.8
3.4
6.8
14.2
32
Summary of Contents for EPM E4418A
Page 20: ...HP4402 book HP4402LOF doc 6 Tue Apr 22 15 20 23 1997 Contents 6 HP EPM 441A User s Guide ...
Page 22: ...HP4402 book HP4402LOT doc 8 Tue Apr 22 15 20 23 1997 Contents 8 HP EPM 441A User s Guide ...
Page 23: ...HP4402 book chapter1 fm 1 Tue Apr 22 15 20 23 1997 1 Getting Started ...
Page 41: ...HP4402 book chapter2 fm 1 Tue Apr 22 15 20 23 1997 2 Power Meter Operation ...
Page 101: ...HP4402 book chapter3 fm 1 Tue Apr 22 15 20 23 1997 3 Menu Reference ...
Page 127: ...HP4402 book chapter4 fm 1 Tue Apr 22 15 20 23 1997 4 Error Messages ...
Page 137: ...HP4402 book chapt5_2 fm 1 Tue Apr 22 15 20 23 1997 5 Specifications ...
Page 158: ...HP4402 book HP4402IX doc 6 Tue Apr 22 15 20 23 1997 Index 6 HP EPM 441A User s Guide ...