HP4402.book : chapter2.fm 29 Tue Apr 22 15:20:23 1997
Power Meter Operation
Measuring Pulsed Signals
HP EPM-441A User’s Guide
2-29
To set the duty cycle:
1.
Press
,
. The current setting of the duty
cycle is displayed under the
softkey.
2.
To change this setting press
. The power meter
displays the duty cycle in a pop up window. Modify this value (see
below) until the desired duty cycle is displayed.
■
Use
or
to modify the digit on which the cursor is
currently positioned.
■
Use
or
to move to other digits.
3.
To confirm your choice press
.
4.
is automatically set to “
On
” when a value is
entered using
.
Duty cycle can be disabled and re-enabled simply by pressing
.
Note
Pulse power averages out any aberrations in the pulse such as
overshooting or ringing. For this reason it is called pulse power
and not peak power or peak pulse power.
In order to ensure accurate pulse power readings, the input signal
must be pulsed with a rectangular pulse. Other pulse shapes (such
as triangle, chirp or Gaussian) will cause erroneous results.
The pulse power on/off ratio must be much greater than the duty
cycle ratio.
System
Inputs
Input Settings
Duty Cycle
Duty Cycle
%
Duty Cycle Off On
Duty Cycle
Duty Cycle Off On
Summary of Contents for EPM E4418A
Page 20: ...HP4402 book HP4402LOF doc 6 Tue Apr 22 15 20 23 1997 Contents 6 HP EPM 441A User s Guide ...
Page 22: ...HP4402 book HP4402LOT doc 8 Tue Apr 22 15 20 23 1997 Contents 8 HP EPM 441A User s Guide ...
Page 23: ...HP4402 book chapter1 fm 1 Tue Apr 22 15 20 23 1997 1 Getting Started ...
Page 41: ...HP4402 book chapter2 fm 1 Tue Apr 22 15 20 23 1997 2 Power Meter Operation ...
Page 101: ...HP4402 book chapter3 fm 1 Tue Apr 22 15 20 23 1997 3 Menu Reference ...
Page 127: ...HP4402 book chapter4 fm 1 Tue Apr 22 15 20 23 1997 4 Error Messages ...
Page 137: ...HP4402 book chapt5_2 fm 1 Tue Apr 22 15 20 23 1997 5 Specifications ...
Page 158: ...HP4402 book HP4402IX doc 6 Tue Apr 22 15 20 23 1997 Index 6 HP EPM 441A User s Guide ...