Guidance and manufacturer’s declaration—electromagnetic immunity
Immunity test
IEC 60601-1-2:2014 home healthcare
test level
Compliance level
Compliance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2, 4, 8, 15 air
± 8 kV contact
± 2, 4, 8, 15 air
Compliant
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV, 100 kHz for power supply lines
± 1 kV, 100 kHz for input/output lines
± 2 kV, 100 kHz for power supply lines
± 1 kV, 100 kHz for input/output lines
Compliant
Surge
IEC 61000-4-5
± 0.5 kv, ± 1 kv line(s) to line(s)
±0.5 kv, ±1 kv, ± 2 kV line(s) to earth
± 0.5 kv, ± 1 kv line(s) to line(s)
±0.5 kv, ±1 kv, ± 2 kV line(s) to earth
Compliant
Voltage dips, short
interruptions, and voltage
variations on power
supply input lines
IEC 61000-4-11
0% U
T
(100 % dip in U
T
) for 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°,
and 315°
70 % U
T
(30 % dip in U
T
) for 25/30 cycles
0% U
T
(100 % dip in U
T
) for 250/300
cycle
0% U
T
(100 % dip in U
T
) for 1 cycle
0% U
T
(100 % dip in U
T
) for 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°,
270°, and 315°
70 % U
T
(30 % dip in U
T
) for 25/30
cycles
0% U
T
(100 % dip in U
T
) for 250/300
cycle
0% U
T
(100 % dip in U
T
) for 1 cycle
40 % U
T
(60 % dip in U
T
) for 5 cycles
Compliant
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
50 Hz and 60 Hz
30 A/m
50 Hz and 60 Hz
Compliant
NOTE:
U
T
is the AC mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration—electromagnetic immunity
The HP ElitePad 1000 G2 Healthcare Tablet is intended for use in the electromagnetic environment specified below. The customer or
the user of the HP ElitePad 1000 G2 Healthcare Tablet should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2:2014 home healthcare
test level
Compliance level
Electromagnetic
environment—
guidance
Conducted RF
IEC 61000-4-6
Applies to both power supply lines and
input/output lines:
3 V
0.15–80 MHz
6 V in ISM and amateur radio bands
a
between 0.15–80 MHz
80% AM at 1 kHz
Applies to both power supply lines
and input/output lines:
3 V
0.15–80 MHz
6 V in ISM and amateur radio bands
a
between 0.15–80 MHz
80% AM at 1 kHz
Compliant
Radiated RF EM Fields
IEC 61000-4-3
10 V/m
80 MHz–2.7 GHz
80% AM at 1 kHz
10 V/m
80 MHz–2.7 GHz
80% AM at 1 kHz
Compliant
IEC 60601-1-2: 2014
67