Guidance and manufacturer’s declaration—electromagnetic immunity
± 0.5 kV, ± 1 kV, ± 2 kV
line(s) to earth
Voltage dips, short
interruptions, and voltage
variations on power
supply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
) for
0.5 cycle
40 % U
T
(60 % dip in U
T
) for 5
cycles
70 % U
T
(30 % dip in U
T
) for 25
cycles
<5 % U
T
(>95 % dip in U
T
) for 5
s
0% U
T
(100 % dip in U
T
) for
0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
40 % U
T
(60 % dip in U
T
) for
5 cycles
70 % U
T
(30 % dip in U
T
) for
25/30 cycles
0% U
T
(100 % dip in U
T
) for
250/300 cycle
0% U
T
(100 % dip in U
T
) for
1 cycle
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the HP
ElitePad 1000 G2 Healthcare Tablet
requires continued operation during
power mains interruptions, it is
recommended that the HP ElitePad 1000
G2 Healthcare Tablet be powered from an
uninterruptible power supply or battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
30 A/m
50 Hz and 60 Hz
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE:
U
T
is the AC mains voltage prior to application of the test level.
Because there can be many different applications/uses of the tablet, the Essential Performance is not
determined at this time and will be defined by the end user.
Guidance and manufacturer’s declaration—electromagnetic immunity
The HP ElitePad 1000 G2 Healthcare Tablet is intended for use in the electromagnetic environment specified below. The customer or
the user of HP ElitePad 1000 G2 Healthcare Tablet should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment—guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
Portable and mobile RF communications equipment
should be used no closer to any part of the HP ElitePad
1000 G2 Healthcare Tablet, including cables, than the
recommended separation distance calculated from the
equipment applicable to the frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz:
800 MHz to 2.5 GHz:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
10 V/m
62
Chapter 12 Specifications