Table 12-2
Ports and cabling – Docked (continued)
Port name on
EUT
Cable description
(or reason for no
cables)
Qty
Length as
tested (m)
Max length
(m)
Shielded?
(Y/N)
Termination box ID &
port name
LAN
Ethernet
1
2.0
>3.0
No
Switch
VGA
VGA
1
1.0
>3.0
No
Monitor
Electromagnetic emissions
Guidance and manufacturer’s declaration—electromagnetic emissions
The HP ElitePad 1000 G2 Healthcare Tablet is intended for use in the electromagnetic environment specified below. The customer or
the user of the HP ElitePad 1000 G2 Healthcare Tablet should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment—guidance
RF emissions
CISPR 11
Group 1
The ElitePad 1000 G2 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The ElitePad 1000 G2 is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Class Complies
WARNING!
Use of this equipment adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the other equipment should
be observed to verify that they are operating normally.
Guidance and manufacturer’s declaration—electromagnetic immunity
The HP ElitePad 1000 G2 Healthcare Tablet is intended for use in the electromagnetic environment specified below. The customer or
the user of the HP ElitePad 1000 G2 Healthcare Tablet should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment—
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV contact
± 2, 4, 8, 15 air
Floors should be wood, concreate, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 0.5 kV, ± 1 kV line(s) to
line(s)
Mains power quality should be that of a
typical commercial or hospital
environment.
EN 60601-1-2: 2007
61