Adjustments
Power Reference Oscillator Level Adjustment
HP EPM-441A/442A Service Guide
3-11
Measurement Uncertainty
When calculating measurement uncertainty the significant multimeter
uncertainty parameters are for V
comp
, V
1
-V
0
and R. Other uncertainties
which should be included are:
•
Maths Assumptions:
±
0.01%
•
Effective Efficiency Calibration:
±
0.05%
•
Mismatch Uncertainty with Source and Mount:
±
0.1%
•
SWR:
≤1.05
These values are typical.