Specifications
Power Meter Specifications
HP EPM-441A/442A Service Guide
1-3
Power Meter Specifications
Meter
Frequency Range
100 kHz to 50 GHz and 75 GHz to 110 GHz, power sensor dependent
Power Range
-70 dBm to +44 dBm (100 pW to 25 W), power sensor dependent
Power Sensors
Compatible with all HP 8480 series power sensors and HP ECP-series
power sensors.
Single Sensor Dynamic Range
90 dB maximum (HP ECP-series power sensors)
50 dB maximum (HP 8480 series power sensors)
Display Units
Absolute:
Watts or dBm
Relative:
Percent or dB
Display Resolution
Selectable resolution of:
1.0, 0.1, 0.01 and 0.001 dB in logarithmic mode, or
1, 2, 3 and 4 significant digits in linear mode
Default Resolution
0.01 dB in logarithmic mode
3 digits in linear mode
4402serv.book Page 3 Monday, March 11, 2002 11:34 AM
Summary of Contents for E4418A
Page 20: ...Contents 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...
Page 22: ...Contents 6 HP EPM 441A 442A Service Guide 4402serv book Page 6 Monday March 11 2002 11 34 AM ...
Page 24: ...Contents 8 HP EPM 441A 442A Service Guide 4402serv book Page 8 Monday March 11 2002 11 34 AM ...
Page 25: ...1 Specifications 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 41: ...2 Performance Tests 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 59: ...3 Adjustments 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 71: ...4 Theory of Operation 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 85: ...5 Replaceable Parts 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 108: ...Index 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...