Specifications
Power Meter Supplemental Characteristics
1-10
HP EPM-441A/442A Service Guide
For HP ECP-series power sensors
In FAST mode, within the range -50 dBm to +17 dBm, for a 10 dB
decreasing power step, the settling time is 10 ms
1
for the HP EPM-441A
and 20 ms
1
for the HP EPM-442A
1
When a decreasing power step crosses the power sensor’s auto-range
switch point, add 25 ms. Refer to the relevant power sensor manual for
further information.
For HP ECP-series power sensors in normal and
x
2 speed modes, manual
filter, 10 dB decreasing power step:
Table 1-5: Settling Time
Number of
Averages
1
2
4
8
16
32
64
128
256
512
1024
Settling Time (s)
(Normal Mode)
0.07
0.12
0.21
0.4
1
1.8
3.3
6.5
13
27
57
Response Time (s)
(x2 mode)
0.04
0.07
0.12
0.21
0.4
1
1.8
3.4
6.8
14.2
32
4402serv.book Page 10 Monday, March 11, 2002 11:34 AM
Summary of Contents for E4418A
Page 20: ...Contents 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...
Page 22: ...Contents 6 HP EPM 441A 442A Service Guide 4402serv book Page 6 Monday March 11 2002 11 34 AM ...
Page 24: ...Contents 8 HP EPM 441A 442A Service Guide 4402serv book Page 8 Monday March 11 2002 11 34 AM ...
Page 25: ...1 Specifications 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 41: ...2 Performance Tests 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 59: ...3 Adjustments 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 71: ...4 Theory of Operation 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 85: ...5 Replaceable Parts 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 108: ...Index 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...