Adjustments
Power Reference Oscillator Frequency Adjustment
3-4
HP EPM-441A/442A Service Guide
Power Reference Oscillator Frequency Adjustment
Note
Adjustment of the power reference oscillator frequency may also
affect the output level of the oscillator. Therefore, after the
frequency is adjusted to 50.0
±
0.5 MHz, the output level should be
checked as described in “Power Reference Oscillator Level
Adjustment”, on page 3-6.
The power reference oscillator frequency is a nominal
specification.
Description
Variable inductor A2L9 is adjusted to set the power reference oscillator
output frequency to 50.0
±
0.5 MHz. This frequency is a supplemental
characteristic.
Figure 3-1: Adjustment Setup
Equipment
Frequency Counter................................ HP 5314A
Procedure
1.
Ensure that the power meter has been powered on for at least
30 minutes before making any measurements.
2.
Connect the equipment as shown in Figure 3-1. Set up the counter
to measure frequency.
Frequency Counter
Power Meter
POWER
REF
4402serv.book Page 4 Monday, March 11, 2002 11:34 AM
Summary of Contents for E4418A
Page 20: ...Contents 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...
Page 22: ...Contents 6 HP EPM 441A 442A Service Guide 4402serv book Page 6 Monday March 11 2002 11 34 AM ...
Page 24: ...Contents 8 HP EPM 441A 442A Service Guide 4402serv book Page 8 Monday March 11 2002 11 34 AM ...
Page 25: ...1 Specifications 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 41: ...2 Performance Tests 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 59: ...3 Adjustments 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 71: ...4 Theory of Operation 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 85: ...5 Replaceable Parts 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Page 108: ...Index 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...