Test 4-1: DC Voltage Test (Zero Volt Input)
This test verifies DC Voltage accuracy on all five ranges using a zero
volt input.
Equipment Setup
1.
Connect the equipment as shown in Figure 4-1.
2.
Set the HP E1411 multimeter as follows:
•
Reset Multimeter ..................................................*RST
•
Auto Zero .................................................................ON
•
Power Line Cycles (PLC) ............................................1
•
Line Freq Reference (CAL:LFR) ...... 50Hz or 60Hz
NOTE
*RST sets Auto Zero to ON and Power Line Cycles to 1.
Test Procedure
Figure 4-1. DC Voltage (Zero Volt Input) Setup
22 Verification Tests
Chapter 4