E1411
Range
10%
overrange
AC Standard Output
Voltage (Vac)
Frequency (Hz)
79.5 mV
79.5 mV
79.5 mV
300.0 V
87.5 mV
87.5 mV
87.5 mV
None
0.07 V
0.07 V
0.07 V
300.00 V
60 Hz
5 kHz
10 kHz
5 kHz
5.
Remove power and disconnect test equipment.
Example: AC Voltage
Test
This example performs an ACV test for a power line reference frequency of
60 Hz. Change line 80 to OUTPUT 70903;"CAL:LFR 50" for 50 Hz
operation.
NOTE
When connected to the HP E1411 multimeter, some AC Standards may
exhibit voltage variations at the start of a measurement. The WAIT 1
statement (line 160) allows settling for the AC Standard output.
1 ! AC Voltage Performance Verification
10 OPTION BASE 1
20 DIM Read_meas(5),Source_volts(5),Source_freq(5)
30 DATA 0.07, 0.07, 0.07, 0.07, 300.0
40 READ Source_volts(*)
50 DATA 20, 60, 5000, 10000, 5000
60 READ Source_freq(*)
70 OUTPUT 70903;"*RST"
!Set autozero ON, PLC 1
80 OUTPUT 70903;"CAL:LFR 60"
!Set 60 Hz line ref frequency
90 FOR I=1 TO 5
100 PRINT "1. Set AC Standard output to";Source_volts(I);"Vac"
110 PRINT "2. Set AC Standard frequency to";Source_freq(I);"Hz"
120 PRINT "3. Press Continue when ready"
130 PAUSE
140 CLEAR SCREEN
150 OUTPUT 70903;"CONF:VOLT:AC";Source_volts(I)
160 WAIT 1
!One second settling time
170 OUTPUT 70903;"READ?"
180 ENTER 70903;Read_meas(I)
190 NEXT I
200 FOR I=1 TO 5
210 PRINT "Voltage for";Source_volts(I);"Vac range @";Source_freq(I);
"Hz=";Read_meas(I);"Vac"
220 NEXT I
28 Verification Tests
Chapter 4
Summary of Contents for E1411A
Page 4: ...Notes ...
Page 8: ...Notes 8 HP E1411A B Service Manual ...
Page 10: ...10 HP E1411A E1411B Service Manual ...
Page 36: ...Notes 36 Verification Tests Chapter 4 ...
Page 48: ...Notes 48 Adjustments Chapter 5 ...
Page 51: ...Figure 6 1 HP E1411A Replaceable Parts Chapter 6 Replaceable Parts 51 ...
Page 53: ...Figure 6 2 HP E1411B Replaceable Parts Chapter 6 Replaceable Parts 53 ...
Page 56: ...56 Manual Changes Chapter 7 ...