90 FOR I=1 TO 5
100 PRINT "Set DC Standard to ";Volts(I);"VDC"
110 PRINT "Press Continue when ready"
120 PAUSE
130 CLEAR SCREEN
140 OUTPUT 70903;"CONF:VOLT:DC";Range(I)
!Set DCV, range
150 WAIT 1
!Wait for settling
160 OUTPUT 70903;"READ?"
170 ENTER 70903;Read_meas(I)
!Enter DC voltage
180 NEXT I
190 FOR I=1 TO 5
200 PRINT "Voltage on";Range(I);"V range =";Read_meas(I);"VDC"
210 NEXT I
220 END
Test 4-3: AC Voltage Test
This test verifies AC voltage accuracy on the 87.5 mV and 300 V ranges
using sine wave inputs at
≥
50% of full scale. The input frequency varies
from 20 Hz to 10 kHz. The DC component must be <10% of the AC
component.
NOTE
The DC Voltage Performance test must be performed prior to the AC
Voltage test to check the A/D accuracy on all ranges. If the DC Voltage test
has not been performed, the AC voltage must be checked on all ranges.
Equipment Setup
1.
Connect the equipment as shown in Figure 4-3.
WARNING
The AC Standard (Datron 4708, Option 20) can produce
dangerous voltages which are present on the terminals. Do not
touch the front (or rear) panel terminals unless you are sure no
dangerous voltage is present.
26 Verification Tests
Chapter 4
Summary of Contents for E1411A
Page 4: ...Notes ...
Page 8: ...Notes 8 HP E1411A B Service Manual ...
Page 10: ...10 HP E1411A E1411B Service Manual ...
Page 36: ...Notes 36 Verification Tests Chapter 4 ...
Page 48: ...Notes 48 Adjustments Chapter 5 ...
Page 51: ...Figure 6 1 HP E1411A Replaceable Parts Chapter 6 Replaceable Parts 51 ...
Page 53: ...Figure 6 2 HP E1411B Replaceable Parts Chapter 6 Replaceable Parts 53 ...
Page 56: ...56 Manual Changes Chapter 7 ...