Test 2-2: DC Offset
Description
The purpose of this test is to verify that the AFG meets its specifications for
DC offset accuracy.
Equipment Setup
–
Connect equipment as shown in Figure 2-4
–
Set DMM to DCV, autorange
Test Procedure
1. Reset the AFG:
*RST;*CLS
Reset AFG and clear status registers
2. Set up the AFG to output the waveform defined above:
OUTP:LOAD INF
;
Infinite load
:VOLT 0.001;
Set amplitude
:FUNC USER
Select user waveform
3. Create a user-defined waveform made up of zeros:
LIST:SEGM:SEL A
Select segment name
LIST:SEGM:VOLT
<array>
Segment list
where <array> is an array of 4096 zeros.
Figure 2-4. DC Offset Test Setup