AC Flatness Adjustment Procedure
Description
An AC adjustment is performed on the AFG by reading a series of voltages
output by the AFG, then entering those values back into the AFG. After all
measurements have been completed, new calibration constants are
calculated and stored in non-volatile memory. To ensure accuracy, perform
the AC flatness calibration procedure at one-year intervals.
This procedure uses a firmware routine to adjust the AFG’s AC calibration
constants. The
CALibration[:AC]:BEGin
command starts the AC calibration
sequence and the
CALibration[:AC]:POINt?
command steps the AFG to the
next calibration point. All AFG settings are performed by the firmware
routine.
NOTE
If an error occurs at any time during the procedure, abort (
*RST
) and start over.
Equipment Setup
–
Perform a complete autocalibration on the DMM (unless an
autocal has been performed within the last 24 hours)
–
Follow the power meter manufacturer’s instructions for
performing an autocalibration and correcting for the power sensor
–
Connect the equipment as shown in Figure 3-2
–
Set the power meter to measure RF power in watts
Figure 3-2. Flatness Adjustment Setup (Power Meter)
Chapter 3
Adjustments 69