Test 2-4: AC Flatness
Description
The purpose of this test is to verify that the AFG meets its specifications for
AC accuracy at frequencies other than 1 kHz.
Equipment Setup
–
Connect equipment as shown in Figure 2-6
–
Set DMM to ACV, 10 V range
Test Procedure
1. Reset the AFG:
*RST;*CLS
Reset AFG and clear status registers
2. Set up the AFG to output a 10 dBm sinewave:
FUNC SIN;
:VOLT 10DBM;
Set amplitude
:
OUTP:LOAD 50
50 ohm load
3. Set the AFG output to the reference frequency (1 kHz):
FREQ 1000
Set frequency
INIT:IMM
Initiate waveform
Figure 2-6. AC Flatness Test Setup (DMM)
Chapter 2
Verification Tests 39