![HP E1340A Service Manual Download Page 22](http://html.mh-extra.com/html/hp/e1340a/e1340a_service-manual_160374022.webp)
Functional Verification: Aux In Test
Description
The purpose of this test is to check the Aux In port. The mainframe Pacer
port is used to gate the output.
Test Procedure
1. Reset the AFG:
*RST;*CLS
Reset AFG and clear status registers
2. Set up the equipment as shown in Figure 2-1. Set the scope to
2 V/div, .02 msec/div.
3. Send the following commands to the mainframe to set up the Pacer
port:
*RST
Reset AFG
PULS:COUN INF
Infinite pulse count
PULS:PER 2 S
2 second period
Figure 2-1. Aux In Test Setup
24 Verification Tests
Chapter 2