HP
8753D
P
erformance
T
est
Record
(2
of
18)
For
30
kHz-6
GHz
Analyzers
Hewlett-Pac
k
ard
Company
Mo del
HP
8753D
Option
011
and
Option
006
Rep
ort
Number
Serial
Numb er
Date
7
7
1.
Source
F
requency
Range
and
Accuracy
CW
F
requency
(MHz)
Lo
w
er
Limit
(MHz)
Measured
V
alue
(MHz)
Upp
er
Limit
(MHz)
Measuremen
t
Uncertain t
y
(MHz)
0.3
0.299
997
0.300
003
6
0.000
000
520
5.0
4.999
950
5.000
050
6
0.000
008
610
16.0
15.999
840
16.000
160
6
0.000
028
220
31.0
30.999
690
31.000
310
6
0.000
053
730
60.999
999
60.999
390
61.000
610
6
0.000
104
800
121.0
120.998
790
121.001
210
6
0.000
206
800
180.0
179.998
200
180.001
800
6
0.000
307
200
310.0
309.995
900
310.003
100
6
0.000
528
300
700.0
699.930
000
700.007
000
6
0.001
191
700
1
300.0
1
299.987
1
300.013
6
0.002
212
300
2
000.0
1
999.980
2
000.020
6
0.003
403
000
3
000.0
2
999.970
3
000.030
6
0.005
104
000
4
000.0
3
999.960
4
000.040
6
0.006
805
000
5
000.0
4
999.950
5
000.050
6
0.008
506
000
6
000.0
5
999.940
6
000.060
6
0.010
207
000
DRAFT
3/21/106
15:21
P
erformance
T
est
Record
2b-3
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com