DISK
MESSAGE
LENGTH
ERROR
Error
Num
b er
190
The
analyzer
and
the
external
disk
driv
e
aren't
communicating
prop erly
.
Chec
k
the
HP-IB
connection
and
then
try
substituting
another
disk
driv
e
to
isolate
the
problem
instrumen
t.
DISK:
not
on,
not
connected,
wrong
addrs
Error
Num
b er
38
The
disk
cannot
b e
accessed
b
y
the
analyzer.
V
erify
p o
w
er
to
the
disk
driv
e,
and
c
hec
k
the
HP-IB
connection
b et
w
een
the
analyzer
and
the
disk
driv
e.
Ensure
that
the
disk
driv
e
address
recognized
b
y
the
analyzer
matc
hes
the
HP-IB
address
set
on
the
disk
driv
e
itself.
DISK
READ/WRITE
ERROR
Error
Num
b er
189
There
may
b e
a
problem
with
y
our
disk.
T
ry
a
new
opp
y
disk.
Ifa
new
opp
y
disk
do es
not
eliminate
the
error,
suspect
hardw
are
problems.
EXCEEDED
7
STANDARDS
PER
CLASS
Error
Num
b er
72
When
mo
difying
calibration
kits,
y
ou
can
dene
a
maxim
um
of
sev
en
standards
for
an
y
class.
INITIALIZATION
FAILED
Error
Num
b er
47
The
disk
initialization
failed,
probably
b ecause
the
disk
is
damaged.
INSUFFICIENT
MEMORY
Error
Num
b er
51
Y
our
last
fron
t
panel
or
HP-IB
request
could
not
b e
implemented
due
to
insucien
t
memory
space.
In
some
cases,
this
is
a
fatal
error
from
whic
h
y
ou
can
escape
only
b
y
presetting
the
instrumen
t.
10-44
Service
Key
Menus
and
Error
Messages