No
de
17
1st
IF
(IF
used
for
phase
lo
c
k)
P
erform
step
A11,
ab o
v
e,
and
then
press
4
MEAS
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ANALOG
IN
4
17
5
4
x1
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COUNTER:
ANALOG
BUS
4
MENU
5
NNNNNNNNNNNNNNNNNNNNNNN
CW
FREQ
.
V
ary
the
frequency
and
compare
the
results
to
the
table
b elow.
Entered
F
requency
Coun
ter
Reading
0.3
to
15.999
MHz
same
as
entered
16
MHz
to
3
GHz
1
MHz
This
no de
displays
the
IF
frequency
(see
no de17)
as
it
en
ters
the
A11
phase
lo
c
k
assem
bly
via
the
A7
ALC
assem
bly
.
This
signal
comes
from
the
R
sampler
output
and
is
used
to
phase
lo
c
k
the
source.
Figure
NODE17
here.
Figure
10-12.
Counter
Readout
Location
10-30
Service
Key
Menus
and
Error
Messages
DRAFT
3/21/106
15:14
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com