A4,
A5,
A6
Samplers
and
A7
Pulse
Generator
T
ools
Required
T-10
TORX
screwdriver
5/16-inc
h
op en-end
torque
wrenc
h
(set
to
10
in-lb)
ESD
(electrostatic
disc
harge)
grounding
wrist
strap
Remo
v
al
1.
Disconnect
the
p o
w
er
cord
and
remov
e
the
top
co
v
er
(refer
to
\Cov
ers"
in
this
c
hapter).
2.
T
o
remov
e
the
R
sampler
(A4),
y
ou
m
ust
remov
e
the
source
brac
k
et
(item
1).
3.
Disconnect
all
cables
from
the
top
of
the
sampler
(A4/A5/A6)
or
pulse
generator
(A7).
4.
Remo
v
e
the
screws
from
the
top
of
the
assem
bly
.
Extract
the
assem
bly
from
the
slot.
Note
If
y
ou
are
removing
the
pulse
generator
(A7),
the
grounding
clip,
whic
h
rests
on
top
of
the
assem
bly
,
will
b ecome
lo
ose
once
the
four
screws
are
remov
ed.
Be
sure
to
replace
the
grounding
clip
when
reinstalling
the
pulse
generator
assem
bly
.
If
y
ou're
removing
more
than
one
sampler,
b e
careful
not
to
mix
them.
The
R
sampler
(A4)
is
dieren
t
from
the
A
and
B
samplers
(A5
and
A6).
Replacement
Rev
erse
the
order
of
the
remov
al
pro cedure.
Note
When
reconnecting
semi-rigid
cables,
it
is
recommended
that
the
connections
b e
torqued
to
10
in-lb.
Be
sure
to
route
W8
and
W9
as
sho
wn.
No
excess
wire
should
b e
hanging
in
the
A11
and
A14
b oard
slots.
Routing
the
wires
in
this
manner
will
reduce
noise
and
crosstalk.
14-26
Assembly
Replacement
and
P
ost-Repair
Procedures
DRAFT
3/21/106
15:15