Installation Guide
Correcting sample vial delivery problems

24

Correcting sample vial delivery problems

When you find a mishandled sample vial, check:

Are there folds or wrinkles in the crimp cap, especially near the neck of the 

sample vial? For more information, see your 

Operation Guide

 and your 

Sampling Techniques

 

Handbook

.

Are you using the correct sample vials? Is the distance from the vial bottom 

to the top of the neck about 28.4 mm? Is the bottle neck diameter 

8.2 ± 0.3 mm? For more information, see your 

Operation Guide

.

If you use labels on the sample vials, are they the correct size? For more 

information, see your 

Operation Guide

.

If you use labels on the sample vials, are they interfering with the gripper? 

For more information, see your 

Operation Guide

.

Are there any obstacles to the tray arm or injector turret motions? Remove 

any obstacles.

Are the tray quadrants and turret in good condition? Clean any residue from 

the sample positions. Temperature-controlled liquid pumped through the 

quadrants, may cause quadrant distortion over time. See your 

Operation 

Guide

.

Summary of Contents for 7683

Page 1: ...Installation Guide HP 7683 Automatic Liquid Sampler ...

Page 2: ...e customer should not attempt to replace the battery or fuses in this instrument Safety Symbols This manual contains safety information that should be followed by the user to ensure safe operation WARNING A warning calls attention to a condition or possible situation that could cause injury to the user CAUTION A caution calls attention to a condition or possible situation that could damage or dest...

Page 3: ...yringe problems 23 Correcting sample vial delivery problems 24 Faults 25 No light is on 26 The Fault light is on 26 The Fault and the Run lights are on 26 Align Mode light is on 27 All lights are on 27 Error messages 28 Bottle in gripper 28 Front or back door open or injector not mounted 28 Front or back injector com error 29 Front or back injector incomplete injection 29 Front or back injector re...

Page 4: ...ack the sampler components Install the ALS Interface Board if purchased Install the tray if purchased See page 2 Install the injector mounting post See page 4 Install the injector See page 4 Install the 8 sample position turret if desired See page 16 Install the bar code reader if purchased See page 6 Connect the cables See page 8 When using the injector for cool on column injection a Check that t...

Page 5: ...vided with the accessory for instructions After it is installed continue with the steps below to install the optional HP G2614A Tray 1 Unpack the injector module tray and bar code reader and place them on the bench top 2 Remove the 3 injection port cover Torx screws shown Figure 1 Installing the tray Injection port cover screws Tray Captive screw install first Tray mounting Remove screws from cove...

Page 6: ...r the 3 holes in the injection port cover and locate in place Tighten the captive screw in the center of the tray mounting arm to hold the tray in place Install the 2 mounting screws into the tray and tighten until the tray is secure 4 Install the tray quadrants Slide the leading edge of each quadrant under the guides on the tray then snap the front tab into the slot on the tray See Figure 2 below...

Page 7: ... on the injection port cover in the front or rear location as desired 2 Open the injector door Slide the packing material and syringe carriage down then remove the packing 3 Mount the injector Line up the hole in the base of the injector that is nearest the cable with the mounting post Lower the injector about an inch 2 5 cm onto the post Figure 3 Mounting the injector Alignment hole Alignment pin...

Page 8: ... will not sit upright on the GC check that the plumbing and cabling under the injection port cover are properly routed in their channels Choosing the turret type to use Your HP G2613A Injector comes with 2 types of turrets a standard turret for use with or without a tray and an alternate 8 sample position turret which is used without a tray If you are using the optional HP G2614A Tray the injector...

Page 9: ...loose You will tighten them later Figure 4 The Bar Code Reader support bracket 2 Mount the Bar Code Reader onto the front of the tray base by inserting the two mounting clips on the Bar Code Reader into the mounting slots on the edge of the tray base See Figure 5 Figure 5 Mounting location for the Bar Code Reader Adjust this space Mounting screws 2 Height adjustment screws Tray Support Bracket GC ...

Page 10: ... the top of the bar code reader is parallel with the tray surface Tighten the mounting screws 5 Adjust the Bar Code Reader s height Slightly loosen the height adjustment screw on the side of the support bracket and raise or lower the bracket until there is a very small gap about the thickness of a piece of paper between the mounting clips on the bar code reader and the tray See Figure 4 Tighten th...

Page 11: ...Automatic Liquid Sampler cabling 2 When the cables are connected turn on the GC power The tray will beep twice After the start up process ends the Ready light on the injector tower should be on If the Align Mode light is on see Turret alignment on page 16 If either the Fault or Run light is on see Faults on page 25 HP 6890 Plus GC G2613A Injector Front G2613A Injector Back Injector cables Tray cab...

Page 12: ... your HP Chemical Analysis Columns and Suppliescatalog and Volume2ofyour GC Operating Manual for a list of parts 2 Prepare the GC inlet See Volume 2 of your GC Operating Manual for instructions Check your needle to column size Verify that the insert matches the needle size If necessary replace the septum 3 Verify that the correct needle support assembly is installed in the injector Replace if nece...

Page 13: ...If necessary reinstall the injector onto the GC 5 Install the syringe See your Operation Guide 6 Rotate the turret clockwise until it stops then verify the installation by manually sliding the syringe carriage down until the needle enters the inlet 530 µm standard 250 µm 320 µm HP accessory 18599T ...

Page 14: ...ure your HP 6890 GC for use with the HP 7683 Automatic Liquid Sampler Refer to the section on injector configuration in your GC Operating Manual for instructions to set the following front and or back injector location nanoliter enable tray enable bar code reader position position 3 enable and type ...

Page 15: ...ample 1 position if not using the tray 3 Set these following parameters on the GC 4 Set the GC oven program to 25 C with a 0 C min ramp a hold time of 0 1 minutes an equilibrium time of 0 3 minutes and an initial time of 0 3 minutes Parameter Setting Injection volume 1 Sample pumps 1 Viscosity delay 0 Sample washes 1 Solve A washes 1 Solv B washes 0 Slow plunger off Pre dwell time 0 Post dwell tim...

Page 16: ...n the sequence If there are no faults the injector will make 1 injection from the first vial position If any faults occur see Faults on page 25 Error messages on page 28 Correcting syringe problems on page 23 or Correcting sample vial delivery problems on page 24 ...

Page 17: ...On an occasional basis Clean the surface of the tray arm gripper gripper jaws and tray quadrants Clean the needle guide and nearby surfaces of the injector Dust and dirt accumulate in these areas and can be picked up by the syringe needle and carried into the inlet Clean the surface of the injector Vacuum off any dust that builds up around the vents Make sure the injector mounting post is tight Ma...

Page 18: ...urret align the tab on the motor hub with the slot in the underside of the turret 5 Push the stripper arm to the back of the tower insert the turret with the open section facing the back of the tower and the arrow pointing back and install the turret with the tab seated in the slot Seat the turret on the hub rotating it gently if necessary until it drops completely into position The top of the thr...

Page 19: ...his procedure has been performed 1 When the injector is not operating open the injector tower door 2 Slide the syringe carriage up until it stops 3 Remove the syringe See your Operation Guide for details 4 Carefully remove the needle support assembly from the injector See Replacing the needle support assembly on page 18 for details 5 Use a pen to press the recessed alignment button above the indic...

Page 20: ...e moves all the way down then moves back up again and clears the turret The syringe carriage steps down until it touches the turret This sets its position relative to the turret The plunger moves to calibrate stops 7 When the alignment is complete the Ready light turns on and the sampler returns to the ready state Install the needle support assembly and syringe If the injector fails alignment see ...

Page 21: ...ections you must change the needle support assembly to HP accessory 18599T Figure 11 Needle support assemblies To replace the needle support assembly 1 Remove all vials from turret and disconnect the injector cable from the GC and lay the HP G2613A Injector on its back on a flat surface 2 Open the injector door 3 Remove the syringe 530 µm 250 µm 320 µm standard HP accessory 18599T 230 µm 320 µm Ne...

Page 22: ... assembly pull gently to release the bearing from the bearing clip in the syringe carriage 5 Carefully use the bearing to pull the rod down until you can lift the assembly out of the syringe carriage Caution Be careful not to pull the assembly by its metal shaft The shaft is easily bent Bearing clip Guide Syringe Slide Bearing Syringe carriage Plunger carrier loop carriage tracks ...

Page 23: ...e carriage tracks as shown in Figure 13 Figure 13 Installing the needle support assembly 250 320 µm shown 8 Align the bearing on the needle support assembly with the plastic bearing clip to the right of the syringe latch and push gently on the bearing until the assembly snaps into place 9 Place the HP G2613A Injector back on the GC or on a parking post Syringe latch Guide Syringe Slide Slide Slide...

Page 24: ...e needle guide in the support foot The needle should slide smoothly in the needle guide Replacing the needle guide in the needle support foot When usingtheHP accessory 18599T for 250 320 µm coolon columninjections the needle guide is replaceable Check the needle guide in the needle support foot for wear whenever you change the inlet septum When the needle guide shows signs of wear replace it as de...

Page 25: ...Installation Guide Replacing the needle guide in the needle support foot 22 Figure 14 Needle support foot 250 320 µm assembly shown Screw Metal plate Needle guide ...

Page 26: ... about 126 5 mm For more information see your Operation Guide and your Sampling Techniques Handbook Are the needle support and needle guide cone clean Remove any residue or septum deposits For more information see Maintenance on page 14 Is the correct insert for your syringe installed in the cool on column inlet For more information see your GC Operation Manual Is the GC septum nut too tight For m...

Page 27: ... 4 mm Is the bottle neck diameter 8 2 0 3 mm For more information see your Operation Guide If you use labels on the sample vials are they the correct size For more information see your Operation Guide If you use labels on the sample vials are they interfering with the gripper For more information see your Operation Guide Are there any obstacles to the tray arm or injector turret motions Remove any...

Page 28: ...ation the Ready light is on If the injector is busy the Run light is on If another combination of lights are on an error has occurred Use the following instructions to try to solve the problem before obtaining HP service Figure 15 Injector status lights Align Mode light Fault light Run light Ready light ...

Page 29: ...Fault light is on Probable causes The injector door is open Suggested actions 1 Ensure that the injector door is closed 2 If the fault light stays on obtain HP service The Fault and the Run lights are on Probable causes The injector is mounted incorrectly on the mounting post Incorrect mounting post Suggested actions 1 Ensure that the injector is mounted properly For more information see page 4 2 ...

Page 30: ... action 1 Verify the turret is properly installed See Removing the turret on page 15 2 Perform the alignment procedure to initalize the system See Turret alignment on page 16 All lights are on Probable cause The cables connections are loose There is a board failure There is a firmware revision conflict Suggested action 1 Check all cable connections 2 Turn the instrument off then on again 3 If the ...

Page 31: ...t delivered properly and stayed in the tray gripper Suggested actions 1 Remove the vial and return it to its position in the tray 2 Ensure that the tray quadrants are snapped into place 3 Ensure that the injector is plugged into the correct connector on the back of the HP 6890 GC and is configured properly See Configuring your GC on page 11 4 Check the deliver to location for the vial and verify t...

Page 32: ... bent The plunger or syringe carriage is operating incorrectly during injection Suggested actions 1 See Correcting syringe problems on page 23 2 Remove the syringe from the injector and check the plunger for stickiness or binding Replace the syringe if necessary 3 Restart the sequence 4 If the error occurs again obtain HP service Front or back injector reset Probable cause There is an interruption...

Page 33: ...the Operation Guide 2 Checktheviscosityofthesampleagainsttheviscosityparameter Reset the viscosity parameter if necessary 3 Restart the sequence 4 If the error occurs again obtain HP service Front or back syringe error Probable causes The syringe carriage motor is defective The syringe is not currently installed or is an incorrect type The syringe carriage sensor is inoperable Suggested actions 1 ...

Page 34: ...ted actions 1 Clear any obstructions 2 Check the Align Mode light If it is lit perform the alignment procedure See Turret alignment in this section 3 Tighten the knurled nut on the top of the turret 4 If the error occurs again obtain HP service Injector not present Probable causes There is a board failure in the injector or GC The injector cable is bad or not securely connected to the GC There is ...

Page 35: ...njector cable is bad or not connected There is a cable failure in the GC Suggested actions 1 Make sure that the injector to GC cable connection is secure 2 If the error remains obtain HP service No bar code reader Probable causes Bar code reader cable not securely connected Bar code reader is defective Tray is defective Suggested actions 1 Ensure the bar code reader cable connection is secure 2 If...

Page 36: ...could not grasp the vial Suggested actions 1 Make sure that the sample vials are in the locations specified by the sequence 2 Make sure that the sample vials meet recommended specification See your Sampling Techniques Handbook 3 If you are using adhesive labels verify that the labels are properly installed See your Operation Guide ...

Page 37: ...ted actions 1 Make sure that the tray cable connection is secure 2 Replace the tray cable 3 If the error remains obtain HP service Tray offline Probable causes There is a board failure in the tray or GC The tray cable is bad or not connected There is a cable failure in the GC Suggested actions 1 Make sure that the tray to GC cable connection is secure 2 Replace the tray cable 3 If the error remain...