6
Operation
In order to remove noise elements, measures should be taken after countermeasures
against shielding, such as surrounding the object to be measured with a metal casing.
If the numerical value still does not stabilize, connect the attached ground wire to the
shield part of the object to be measured, and set the shield part of the target and the
measuring object to the same potential. (3 terminal measurement)
【
Countermeasure
】
If it is difficult to measure in an environment where the noise environment has been
removed, such as on-site daily management, it is necessary to measure (preliminary
evaluation) in a noise-free environment before using the ESD protection product, Please
perform measurement before starting use.
(The measured value will change depending on the influence of humidity, temperature,
etc. Please adjust the measurement conditions as much as possible.) If the humidity
exceeds 60%, measurement results within the accuracy range will not be obtained.
If the measurement results of both are different, it can be inferred that the difference is
an external factor such as a noise factor.
In daily inspection, we recommend that you manage this considering the difference.
【
About daily management
】
Tool
Shield
Insulator
Attached earth
wire
Please insert the earth wire
terminal through the monaural
plug and insert it tightly until
the terminal touches the metal
of the jack.