7

About measurement mode

Measurement mode can be switched by mode switch.

The following constant measurement mode is set at shipment.

<Always measuring mode>

The measured value is displayed at all times.

It automatically turns off in 35 seconds from the start of measurement.

<Measurement value hold mode>

Fixed the measurement value after 15 seconds from the start of measurement, turn OFF 

after displaying for 5 seconds.

When the object to be measured is in the vicinity of the switching point of applied voltage 

(around 1.00 x 10

6

 

Ω

), the applied voltage may go back and forth between 100 V / 10 V 

and the measurement result may not be stabilized.

In this case, please adopt measurement result narrowed down to either applied voltage.

Measurement value In the hold mode, it is sometimes fixed as a result of an unintended 

applied voltage, so we recommend using it in the continuous measurement mode.

How to switch measurement mode

Remove the protective case and remove the 4 screws 

on the back side, the main case opens.

When the switch at the upper right of the battery box 

is tilted to the left, it is always in measurement mode, 

and when it is tilted to the right, it enters measurement 

value hold mode.

After switching, install the main unit case and the 

protective case in reverse order.

When the object to be measured is high resistance (× 10

10

 to 10

11

 

Ω

 or more), it takes 

time to stabilize the measured value.

Because measured values may not be stabilized within 15 seconds, we recommend you 

to use in constant measurement mode when measuring high resistance.

Measurement value When using hold mode, please measure twice or more and adopt 

later measurement result.

The current mode can be confirmed with the 15 S mark.
Always measuring mode ... Not displayed.
Measurement value hold mode ... Displays for 15 seconds from power on.

Display panel - Always measuring mode

 

Selector switch

Always 

Always 

measuring

measuring

Always 

measuring

Hold

Display panel - Measured value hold mode

Summary of Contents for F-109

Page 1: ...efore attempting to place this instrument in service Please keep this instruction manual available for reference INSTRUCTION MANUAL Identification of parts and contents Connection wire white Measureme...

Page 2: ...over less than 2 00 1012 1 00 103 or over less than 1 00 1012 10 20 for 1 00 1011 or higher 1 When less than 1 00 x 106 10 V 5 switched automatically When 1 00 x 106 or over 100 V 5 Alkaline battery...

Page 3: ...unction 5 When the device will not used for an extended period store without the battery 6 For this is a precision device avoid strong impact vibration and high temperature Caution Precautions Prepara...

Page 4: ...cord white and the electrode cord jack for monaural plug with the connection cord gray 1 Connect the electrode Please use the attached electrode alligator clip according to the measurement application...

Page 5: ...When 100 V is applied 3 When you press and hold the measurement switch 0 5 sec or more the resistance value is displayed on the display and measurement starts When the object is less than 1 106 the ap...

Page 6: ...minary evaluation in a noise free environment before using the ESD protection product Please perform measurement before starting use The measured value will change depending on the influence of humidi...

Page 7: ...urement mode How to switch measurement mode Remove the protective case and remove the 4 screws on the back side the main case opens When the switch at the upper right of the battery box is tilted to t...

Page 8: ...n the cord jack and the conductive pad is 400 or less and the resistance of the circuit as shown in the figure is 800 or less About weight Weight 2 3 kg 1 piece conforms to the standard 2 5 kg 0 25 kg...

Reviews: