8
On compatibility of electrode device to standard
Maintenance and calibration
About external dimensions
Although the external dimensions are
79 mm
φ
× 110 mm, the conductive pad
in contact with the object to be measured is
63.5 mm
φ
and conforms to the standard.
On the resistance of the electrode device
RCJS-5-1 stipulates that "the inter-point r
esistance between two probes must have
sufficient conductivity so that it is lower
than 10
3
Ω
".
For our products, the resistance per
electrode (between the cord jack and the
conductive pad) is 400
Ω
or less, and the
resistance of the circuit as shown in the
figure is 800
Ω
or less.
About weight
Weight 2.3 kg (1 piece) conforms to the
standard 2.5 kg ± 0.25 kg.
・
When the electrode becomes dirty please wipe gently with alcohol etc.
Do not use thinners, benzine, etc.
・
We also have calibration certificate attached (F-109-TA, F-109-1-TA).
Please re-calibrate after use for a certain period through our purchase store.
Surface resistance
meter included code
Surface Resistance Checker
63.5 mm
φ
F-101
Metal plate (approximately 0
Ω
)
F-101
Below 10
3
Ω
1-2-12 Saiwaicho, Naniwa-ku, Osaka 556-0021, Japan
Tel : +81-6-6567-3111 Fax : +81-6-6562-0024
19.02
400
Ω
or less 400
Ω
or le
ss