8

On compatibility of electrode device to standard

Maintenance and calibration

About external dimensions

Although the external dimensions are 

79 mm

φ

 × 110 mm, the conductive pad 

in contact with the object to be measured is 

63.5 mm

φ

 and conforms to the standard.

On the resistance of the electrode device

RCJS-5-1 stipulates that "the inter-point r

esistance between two probes must have 

sufficient conductivity so that it is lower 

than 10

3

 

Ω

".

For our products, the resistance per 

electrode (between the cord jack and the 

conductive pad) is 400 

Ω

 or less, and the 

resistance of the circuit as shown in the 

figure is 800 

Ω

 or less.

About weight

Weight 2.3 kg (1 piece) conforms to the 

standard 2.5 kg ± 0.25 kg.

  When the electrode becomes dirty please wipe gently with alcohol etc.

 

Do not use thinners, benzine, etc.

  We also have calibration certificate attached (F-109-TA, F-109-1-TA). 

 

Please re-calibrate after use for a certain period through our purchase store.

Surface resistance 
meter included code

Surface Resistance Checker

63.5 mm

φ

F-101

Metal plate (approximately 0 

Ω

)

F-101

  

 

Below 10

3

 

Ω

1-2-12 Saiwaicho, Naniwa-ku, Osaka 556-0021, Japan
Tel : +81-6-6567-3111   Fax : +81-6-6562-0024

19.02

400 

Ω

 

or less 400

 

Ω

 or le

ss

Summary of Contents for F-109

Page 1: ...efore attempting to place this instrument in service Please keep this instruction manual available for reference INSTRUCTION MANUAL Identification of parts and contents Connection wire white Measureme...

Page 2: ...over less than 2 00 1012 1 00 103 or over less than 1 00 1012 10 20 for 1 00 1011 or higher 1 When less than 1 00 x 106 10 V 5 switched automatically When 1 00 x 106 or over 100 V 5 Alkaline battery...

Page 3: ...unction 5 When the device will not used for an extended period store without the battery 6 For this is a precision device avoid strong impact vibration and high temperature Caution Precautions Prepara...

Page 4: ...cord white and the electrode cord jack for monaural plug with the connection cord gray 1 Connect the electrode Please use the attached electrode alligator clip according to the measurement application...

Page 5: ...When 100 V is applied 3 When you press and hold the measurement switch 0 5 sec or more the resistance value is displayed on the display and measurement starts When the object is less than 1 106 the ap...

Page 6: ...minary evaluation in a noise free environment before using the ESD protection product Please perform measurement before starting use The measured value will change depending on the influence of humidi...

Page 7: ...urement mode How to switch measurement mode Remove the protective case and remove the 4 screws on the back side the main case opens When the switch at the upper right of the battery box is tilted to t...

Page 8: ...n the cord jack and the conductive pad is 400 or less and the resistance of the circuit as shown in the figure is 800 or less About weight Weight 2 3 kg 1 piece conforms to the standard 2 5 kg 0 25 kg...

Reviews: