![horiba NAVI F-51 Instruction Manual Download Page 96](http://html1.mh-extra.com/html/horiba/navi-f-51/navi-f-51_instruction-manual_2162272096.webp)
Chapter 6 Maintenance and Troubleshooting
6.2 Troubleshooting
88
HORIBA
●
Measurements are not repeatable
●
When the printer will not print even though it is connected.
Check the following points:
・
Is the printer turned ON?
・
Has a printer error occurred?
・
Is there printing paper? Has the paper jammed?
・
When running a test print according to the manual, does it print out
correctly?
Cause
How to solve problem
Some effects of the
sample solution
The pH or other properties of the sample
solution may have changed over time, making
repeatability poor.
The responsive
membrane is dry or
dirty.
Wash the responsive membrane.
There is not enough
reference solution or
it is dirty.
Replace the reference solution with new
solution.
The responsive
membrane is
cracked or worn out.
Replace the electrode.