ERR No. 01 Memory error
Explanation
Data cannot be read from or written to the internal memory.
●
ERR No. 02 Battery voltage low
Explanation
The battery has insufficient voltage.
The measured value cannot be guaranteed when ERR No. 02 is displayed.
Cause
How to solve problem
The pH meter does
not start operating
correctly even after
the power is turned
ON.
Take the battery from the pH meter, and
disconnect the AC adapter. Then press the
ON/OFF key for about 10 seconds.
The internal IC is
defective.
Seek repairs at your nearest retail outlet or
HORIBA service station.
Cause
How to solve problem
The battery voltage
is low.
(Battery voltage: 4.4
V or less)
Replace the dry-cell battery.