Chapter 5 Printer
5.4 Printing format
72
HORIBA
5.4.5 When the ENTER key is pressed in the calibration history display
●
pH calibration history (without check data)
●
pH calibration history (with check data)
5.4.6 Test printing format in the Maintenance mode
Electrode status:
GOOD: good condition
CHECK: Washing is needed.
BAD: Replace
Sensitivity display
Standard solution value is dis-
played.
No calibration/1-point calibration:
None
2-point calibration: 1st item
3-point calibration: 2nd item
Asymmetrical poten-
tial at calibration
CALIBRATION ; BAD
Date
; 2003/ 1/ 1
mV Value
; -7mV
Slope
pH 6.86- 4.01 86. 4%
CAL DATA
2003/ 1/ 1 0:19
pH 6.86 0mV
ATC 25.0°C
2003/ 1/ 1 0:19
pH 4.01 146mV
ATC 25.0°C
Calibration
data
Value of repeatability
check
Calibration
data