![horiba NAVI F-51 Instruction Manual Download Page 32](http://html1.mh-extra.com/html/horiba/navi-f-51/navi-f-51_instruction-manual_2162272032.webp)
Chapter 2 Taking Measurements
2.5 Measuring pH
24
HORIBA
4.
Press the MEAS key while in the Instantaneous Value
Measurement screen.
HOLD will blink until the reading stabilizes.
When the indicated value stabilizes, HOLD will stop blinking and
will be displayed. The indicated value will remain displayed
continually.
Refer to the “ Criteria for judging stability” page 14 for the criteria for
judging the stability of the readout.
When measurement data is held using Instantaneous Value Measurement or
Auto Hold Measurement, you can store that data in the memory by pressing the
▲
key. See