Error Messages and Troubleshooting
51
The indicated value does not change/No response
The measured value is out of the measurement range
When the measured value is below the display range, “Ur” appears. When the measured value
is over the display range, “Or” appears.
Some effects of the sample.
Response time may slow down, depending on
the properties of the sample solution.
Cause
Solution
The electrode is broken.
Replace the electrode.
The electrode is not connected
correctly.
Connect the electrode correctly.
Keys are locked.
Turn OFF the power, remove the batteries, and
then turn ON the power again.
The instrument is in HOLD state.
Push the MEAS, and cancel the HOLD state.
Instrument defect.
Consult your dealer.
Cause
Solution
Sample is out of the measurement
range.
Use a sample within the measurement range.
The conductivity electrode cable is
broken.
Replace the electrode.
Cell constant is not set.
Set the cell constant.
Calibration is not performed or
performed incorrectly.
Perform calibration correctly.
Instrument defect.
Check as explained below.
Cause
Solution
Summary of Contents for LAQUA-EC1500
Page 1: ...Instruction Manual Benchtop Conductivity Meter LAQUA EC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...Conductivity Setup 28...
Page 50: ...TDS Setup 32...
Page 64: ...General Setup 46...
Page 76: ...Printout Summary 58 Appendix 2 Printer Format Measurement Conductivity Resistivity Salinity...
Page 77: ...Printout Summary 59 TDS Printer Format Data Log...
Page 78: ...Printout Summary 60 Printer Format Calibration Conductivity...
Page 79: ......