
Conductivity Calibration
12
Auto Conductivity Calibration
1. After placing the conductivity electrode in the standard solution, press the
key.
2. Meter displays “Auto cal” as per set calibration method and starts measuring various
calibration values with a
blinking on screen.
3. Wait for the
to stabilize (stable calibration reading). When it stabilizes, there will be a
sound.
4. Press the
ENT
key to confirm and save calibration data.
5. Meter displays
DONE
indicating end of the conductivity calibration procedure.
6. Repeat for other calibration points as required.
7. You can calibrate one point for each range.
To view and /or print calibration data, press the
key when you are in the
CAL
mode. The display scrolls through the calibrated values and calibration factor (average
calibration factor for multi-point calibration).
key
key
Note
Summary of Contents for LAQUA-EC1500
Page 1: ...Instruction Manual Benchtop Conductivity Meter LAQUA EC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...Conductivity Setup 28...
Page 50: ...TDS Setup 32...
Page 64: ...General Setup 46...
Page 76: ...Printout Summary 58 Appendix 2 Printer Format Measurement Conductivity Resistivity Salinity...
Page 77: ...Printout Summary 59 TDS Printer Format Data Log...
Page 78: ...Printout Summary 60 Printer Format Calibration Conductivity...
Page 79: ......