Conductivity Calibration
13
Manual Conductivity Calibration
1. After placing the conductivity electrode in the standard solution, press the
key.
2. Meter displays “Manual cal” as per set calibration method and starts measuring various
calibration values with a blinking
on screen.
3. Wait for the
to stabilize (stable calibration reading). When it stabilizes, there will be a
sound.
4. Use the
keys to enter the electrical conductivity value of the standard solution used
for calibration.
5. Press the
ENT
key to confirm and save calibration data.
6. Meter displays
DONE
indicating end of the conductivity calibration procedure.
7. Repeat for other calibration points as required.
8. You can calibrate one point for each range.