Error Messages and Troubleshooting
50
● Troubleshooting
This section describes the causes of problems and their solutions that customers frequently
ask.
The indicated value fluctuates
< Problem with the electrode >
<
Problem with the instrument
>
< Problem with the sample >
The response is slow
Cause
Solution
The conductivity electrode is dirty.
Clean the conductivity cell.
The conductivity electrode is broken.
Replace the conductivity cell.
There are air bubbles between the
conductivity electrode sleeve and
metal rings.
Shake the conductivity electrode to remove
the air bubbles.
Cause
Solution
There is a motor or other device causing
electrical interference.
Measure at a place where no influence from
induction is given. Ground all AC-powered
equipment.
The conductivity electrode is not
connected correctly.
Connect the conductivity electrode properly.
Cause
Solution
Some effects of the sample.
Response time may slow down, depending on
the properties of the sample solution.
Cause
Solution
The conductivity electrode is dirty.
Clean the conductivity cell.
The conductivity electrode is broken.
Replace the conductivity cell.
Summary of Contents for LAQUA-EC1500
Page 1: ...Instruction Manual Benchtop Conductivity Meter LAQUA EC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...Conductivity Setup 28...
Page 50: ...TDS Setup 32...
Page 64: ...General Setup 46...
Page 76: ...Printout Summary 58 Appendix 2 Printer Format Measurement Conductivity Resistivity Salinity...
Page 77: ...Printout Summary 59 TDS Printer Format Data Log...
Page 78: ...Printout Summary 60 Printer Format Calibration Conductivity...
Page 79: ......