When an error message appears
90
●
ERR No.0008
Calibration interval error
・
More than the set number of days has elapsed since calibration was last performed.
・
The calibration interval alarm is "ON" and calibration has not been performed.
●
ERR No.0009
Printer error
An error occurred during printer communication.
●
ERR No.0010
Memory full
Attempted to save more than the specified number of items.
Cause
How to solve problem
Calibration has not been
performed for the set number of
days of the calibration interval or
longer.
Perform calibration.
The calibration interval alarm is
"ON" and calibration has not
been performed.
Perform calibration.
Cause
How to solve problem
There is a problem with the
printer unit connection.
Check the printer connection, and connect the instrument
and printer again.
The defect of the printer
Consult your dealer.
Cause
How to solve problem
Saving more than specified
items of data is attempted.
The maximum number of savable items of data is 100
(LAQUAact-PH110) or 1000 (LAQUAact-PH120 and
LAQUAact-PH130). Copy or transfer necessary data to a
PC and delete the data from the memory (“ Deleting all
saved data ” (page 56)).