![Honeywell STT850 Series User Manual Download Page 65](http://html.mh-extra.com/html/honeywell/stt850-series/stt850-series_user-manual_155005065.webp)
Revision 4.0
STT850 Series HART/DE Option User’s Manual
Page 55
5.2.13 Setting Range Values for Applied Temperature
When setting the range values using applied Temperature, the URV changes
automatically to compensate for any changes in the LRV and to maintain the present span
(URV
– LRV). When entering the LRV using the Tookit keypad, the URV does not change
automatically.
If you use the applied temperature method, and need to change the LRV and URV,
change
the LRV first
. You can also use the local zero and span adjustments on the Transmitter to set
the LRV and URV values.
1.
Starting at the
My Device
menu, make the following menu selections:
Shortcuts > Device setup > Calibration > Calibration Methods > Apply values.
2.
Click
Execute
. You will be warned to remove the loop from automatic control. After doing
so, press
OK
to continue.
3.
Select
4mA
from the list, and then click
OK
. A message will prompt you to apply a new
4 mA input.
4.
Click
OK
; otherwise, click
Abort
.
5.
When the
Current applied process value:
is displayed, choose
Select as 4mA value
, and
click
OK
.
6.
Repeat steps 2 through 4 to set the URV to the applied input Temperature for 20 mA output.
7.
Click
Return
to go back to the Calibration menu.
8.
Click
Send
. The Send to Device screen will be displayed.
9.
Select the
Apply Values
check-box.
10.
Click
Send
to download the change to the Transmitter, or click
Return
to continue making
changes.
5.2.14 Saving device history
FDC provides you a feature wherein you can save the device configuration snapshot as history. This
history record may then be transferred to a central asset management database such as FDM.
Using this feature you can save the device configuration snapshot as device history of a connected
device at any given time in a predefined location. The following are the features of save device
history option.
Two formats of history are supported: FDM and DocuMint.
Only one snapshot per device instance is allowed to be saved and you can save the
snapshot of a device any number of times overwriting the existing one.
To save device history, perform the following steps.
1.
On Device Home page, tap Tools.
2.
Select
Save History
and tap
Select
The
Save History
page appears.